Glutathione detection in blood components: Analytical methods, biosensing strategies, and microfluidic opportunities.

Ludovico Migliaccio, Jan Brodský, Jaromír Jarušek, Imrich Gablech, Roman Guráň, Tomáš Do, Ondřej Zítka, Pavel Neužil

Faculty of Electrical Engineering and Communication, Department of Microelectronics, Brno University of Technology, Technická 3058/10, 616 00, Brno, Czech Republic; Department of Chemistry and Biochemistry, Mendel University in Brno, Zemědělská 1, 61300, Brno, Czech Republic. Electronic address: migliaccio@vutbr.cz. · Faculty of Electrical Engineering and Communication, Department of Microelectronics, Brno University of Technology, Technická 3058/10, 616 00, Brno, Czech Republic; Department of Chemistry and Biochemistry, Mendel University in Brno, Zemědělská 1, 61300, Brno, Czech Republic.

Review

Key points

From the paper's conclusion
  • Progress toward routine clinical use will therefore depend less on lower detection limits than on integrated, standardized sample-to-answer systems, for which microfluidic platforms combining blood handling, stabilization, detection and data processing offer a promising, but not yet demonstrated, route.

Abstract

Reliable determination of glutathione (GSH) remains analytically challenging because the analyte oxidizes rapidly after sampling, lacks an intrinsic chromophore or fluorophore, is highly polar, and occurs in matrices, above all whole blood and erythrocytes, that are optically opaque and rich in competing thiols. This review critically compares the analytical strategies developed to meet that challenge, covering spectrophotometry, fluorescence, chromatography, mass spectrometry, immunochemistry, electrochemistry and microfluidic implementation, and evaluates each with respect to analytical performance, sample preparation, matrix compatibility and suitability for point-of-care use. Electrochemical platforms are best suited to point-of-care integration, because their signal is unaffected by optical opacity, but their selectivity against other thiols, resistance to electrode fouling and batch-to-batch reproducibility have been shown mainly in buffer or spiked samples, not in clinical blood; fluorescence, in turn, enables intact-cell and single-cell analysis, while chromatographic and mass spectrometric methods remain laboratory reference standards on account of their superior molecular specificity and isotope-dilution calibration. Preservation of the native redox state of GSH and glutathione disulfide (GSSG) during sampling, lysis and handling, however, remains across all platforms the dominant source of analytical uncertainty, more than detector sensitivity, and it is compounded by the absence of standardized preparation protocols, certified reference materials and matrix-specific validation. Progress toward routine clinical use will therefore depend less on lower detection limits than on integrated, standardized sample-to-answer systems, for which microfluidic platforms combining blood handling, stabilization, detection and data processing offer a promising, but not yet demonstrated, route.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers