Talanta

Automated online microdialysis coupled with capillary electrophoresis via a three-layer PDMS microchip cast in 3D-printed moulds for rapid profiling of free amino acids in milk.

Petr Tůma

Department of Hygiene, Third Faculty of Medicine, Charles University, Ruská 87, Prague 10, 100 00, Czechia. Electronic address: petr.tuma@lf3.cuni.cz.

Key points

From the paper's conclusion
  • A fully automated online analytical platform coupling microdialysis (MD) with capillary electrophoresis and contactless conductivity detection was developed for the rapid, label-free profiling of free amino acids (FAAs) and related small nitrogenous metabolites in complex milk matrices.

Abstract

A fully automated online analytical platform coupling microdialysis (MD) with capillary electrophoresis and contactless conductivity detection was developed for the rapid, label-free profiling of free amino acids (FAAs) and related small nitrogenous metabolites in complex milk matrices. A core technological innovation of this system is the fully automated online interface, realized via a custom three-layer poly (dimethylsiloxane) (PDMS) microfluidic chip fabricated using high-precision 3D-printed master moulds. Operating in fluidic gate injection mode, this robust three-layer microchip seamlessly bridges an external MD probe directly to a commercial CE instrument under complete software control. By running the MD in a pulsed, discontinuous flow regime, sufficient residence time was provided for analyte mass transport to reach thermodynamic equilibrium across the membrane, yielding high recoveries (95.1-98.4 %). Matrix salinity and sample acidification were strategically exploited to induce transient isotachophoresis sample stacking, significantly enhancing electrophoretic separation efficiency (up to 1,445,000 theoretical plates/m). Under optimized conditions (1.0 M acetic acid at pH 2.4 as BGE; 25 μm capillary; -20 kV separation voltage with applied hydrodynamic back-pressure), 25 target metabolites were simultaneously resolved within 10 min. Calibration curves demonstrated excellent linearity (R2 > 0.9994) with LOQs ranging between 0.11 and 0.38 μM in raw sample matrices. The method exhibited high repeatability, with intra-day RSDs of 0.17-0.24 % for migration times and 0.3-0.6 % for peak areas. The fully automated, three-layer PDMS microchip-based CE-MD system effectively eliminated macromolecular interferences and was successfully applied to map distinct FAA signatures across fresh cow's, UHT cow's, goat's, and human milk.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers