Quantitative screening of Legionella pneumophila in preconcentrated water samples using a portable SERS lateral-flow immunoassay.
Pedro Quintero-Campos, Sophia C Bittinger, Karla Chavarri-Urraca, Sara Martorell, Ángel Maquieira, Dorleta Jimenez de Aberasturi, Sergi Morais
Instituto Interuniversitario de Investigación de Reconocimiento Molecular y Desarrollo Tecnológico (IDM), Universitat Politècnica de València-Universitat de València, Valencia, Spain. · CIC biomaGUNE, Basque Research and Technology Alliance (BRTA), 20014, Donostia-San Sebastián, Spain.
Key points
From the paper's conclusion- These results demonstrate that portable SERS readout enables rapid and repeatable L. pneumophila screening in preconcentrated water samples with high relative specificity, supporting sample prioritization while retaining culture-based confirmation for definitive regulatory interpretation.
Abstract
Screening for Legionella pneumophila is needed for timely water-safety management, whereas conventional culture-based surveillance can take several days. Here, we developed a surface-enhanced Raman scattering lateral-flow immunoassay (SERS-LFIA) that combines biphenyl-4-thiol-encoded gold nanostars (AuNSt-BPT), sandwich immunorecognition, and a portable 785-nm Raman spectrometer for the analysis of preconcentrated water samples. One-liter samples were concentrated to 50 mL, and strip processing plus Raman acquisition required approximately 30 min, excluding preconcentration. Analytical performance was established using a weighted quadratic calibration model and evaluated by blinded analysis of 57 naturally contaminated water-sample eluates against culture, followed by PCR confirmation. Each sample result was obtained from three independently processed strips. At the prespecified nominal original-water-equivalent decision threshold of ≥103 CFU/L (≥2.0 × 104 CFU/L in the conditioned eluate), the SERS-LFIA correctly classified 24 of 26 reference-positive samples and all 31 reference-negative samples, yielding 92.3% relative sensitivity, 100% relative specificity, and a Cohen's κ of 0.929. The buffer-derived operational limit of quantification was 3.20 log10 CFU/L (1.6 × 103 culture-equivalent CFU/L in the conditioned eluate), with within-run inter-strip relative standard deviations below 8%. For the 30 samples above this limit, agreement with culture showed a mean SERS-LFIA-minus-culture difference of 0.039 log10 CFU/L and 95% limits of agreement from -0.711 to 0.790 log10 CFU/L. These results demonstrate that portable SERS readout enables rapid and repeatable L. pneumophila screening in preconcentrated water samples with high relative specificity, supporting sample prioritization while retaining culture-based confirmation for definitive regulatory interpretation.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Combination of antibody-based screening methods for the identification and quantification of Legionella pneumophila in process water.Analytical and bioanalytical chemistry · 2026
- LeVoScan; a pre-enrichment PCR method for rapid detection of culturable Legionella pneumophila in water.Journal of microbiological methods · 2026
- Multimodal LFIA platform integrating visual readout, smartphone grayscale imaging and surface-enhanced Raman scattering for point-of-care detection of SARS-CoV-2 protein.Analytical methods : advancing methods and applications · 2026
- Filtration-Enhanced Dual-Mode SERS Biosensor for Rapid On-Tap Detection of Waterborne Pathogens in Drinking Water.Biosensors · 2026
- From Culture to Omics: Laboratory Methods for Environmental Detection of Legionella in Water Systems.Microorganisms · 2026
- 62 Optimization of Legionella pneumophila Diagnostic Testing in a Large Healthcare System in Southeast TexasAmerican Journal of Clinical Pathology · 2026