A Portable UV-LED Dual-Nafion Microfluidic Platform for Label-Free Trace Protein Preconcentration and Absorbance Detection.
Yi-Jia Shih, Yu-Sheng Chen, Chun-Ping Jen
State Key Laboratory of Mariculture Breeding, Key Laboratory of Marine Biotechnology of Fujian Province, College of Marine Sciences, Fujian Agriculture and Forestry University, Fuzhou 350002, China. · Department of Mechanical Engineering and Advanced Institute of Manufacturing for High-Tech Innovations, National Chung Cheng University, Chiayi 62102, Taiwan.
Key points
From the paper's conclusion- Here, we present a portable UV-LED dual-Nafion microfluidic platform that integrates a 280 nm UV-LED, a chromium-coated slot collimator, a dual-Nafion ion concentration polarization (ICP) preconcentrator, and a miniature spectrometer within a 3D-printed enclosure.
Abstract
Sensitive label-free detection of low-abundance proteins requires both effective on-chip preconcentration and a compact, low-noise optical readout for use outside conventional laboratories. Here, we present a portable UV-LED dual-Nafion microfluidic platform that integrates a 280 nm UV-LED, a chromium-coated slot collimator, a dual-Nafion ion concentration polarization (ICP) preconcentrator, and a miniature spectrometer within a 3D-printed enclosure. The symmetric dual-Nafion geometry on a UV-transparent quartz substrate confines the ICP enrichment zone to a fixed 120 × 500 µm interrogation window, enabling spatially stable, label-free on-chip absorbance monitoring at 280 nm. Using bovine serum albumin (BSA) as a model protein, the platform showed a linear absorbance response over 0-60 µM (R2 = 0.9933) and a static-loading limit of detection of 0.91 µM. Under a 30 V bias in 0.3 mM phosphate-buffered saline, representative 60 min experiments produced apparent enrichment factors of approximately 41, 4.5 × 103, and 1.6 × 104 at initial BSA concentrations of 1 µM, 10 nM, and 1 nM, respectively. At 10 pM, a reproducible, voltage-dependent absorbance increase was detected with three independent chips, whereas the 0 V and BSA-free controls remained below the detection threshold. The platform provides a practical approach for label-free trace-protein analysis and a potential upstream module for biomarker-specific assays.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Smartphone-integrated portable microfluidic platform for liver biomarker quantification using deep learningScientific Reports · 2025
- Finger-actuated microfluidic device for point-of-care circulating cell-free nucleic acid extraction.Lab on a chip · 2026
- Biomimetic fluidic lipid-engineered nanofluidic platform for ultrasensitive discrimination of multiprotein in biofluids.Chemical science · 2026
- High-Performance Narrowband Near-Ultraviolet Photodetector with Array-Based Dynamic ImagingACS Nano · 2026
- Bubble-cell native fluorescence imaging for high-sensitivity label-free protein quantification.Talanta · 2026
- Dual-Signal Amplification: Synergistic Integration of Nanozyme-Catalyzed SERS Aptasensor with μHPLC Pre-enrichment for Ultrasensitive ATP Monitoring.Analytical chemistry · 2026