Research and practice in thrombosis and haemostasis

A bicentric pilot study of apixaban and rivaroxaban quantification using a novel point-of-care system in the emergency department.

Cindy Tissier, Tiffany Madranges, Nicolas Peschanski, Véronique Le Meur, Thibault Donnet, Emmanuel De Maistre, Isabelle Gouin-Thibault

Emergency Department, University Hospital of Dijon, Dijon, France. · Hematology and Hemostasis Laboratory, University Hospital of Dijon, Dijon, France.

Observational

Key points

From the paper's conclusion
  • The POC system shows a high level of agreement with a laboratory anti-Xa assay and reduced turnaround time, supporting its potential use in the emergency setting.

Abstract

BACKGROUND: Direct oral anticoagulants (DOACs) do not require routine monitoring; however, rapid assessment of drug concentrations may be needed in critical situations. Laboratory-based anti-Xa assays are recommended but are not consistently available with turnaround times compatible with emergency care. OBJECTIVES: This study evaluated analytical performance, agreement at clinical thresholds, precision, and turnaround time of a whole-blood point-of-care (POC) system for measuring apixaban and rivaroxaban. METHODS: A preliminary laboratory study and a prospective bicentric pilot study in emergency departments were performed on DOAC-treated patients. Paired samples were analyzed using the POC system and a laboratory anti-Xa assay (STA-R Max3; Diagnostica Stago). Agreement within the analytical measuring range (23-200 ng/mL) was assessed using Passing-Bablok regression and Bland-Altman analysis. Concordance at 30, 50, and 100 ng/mL was evaluated using Cohen κ and prevalence- and bias-adjusted κ. Precision and time-to-result were also assessed. RESULTS: Excellent agreement was observed in both the preliminary laboratory study (r = 0.978; slope = 0.994; median bias = -5.5 ng/mL) and the emergency room study in the 69 analyzed patients (r = 0.965; slope = 1.013; median bias = -10.5 ng/mL). Agreement increased with higher cutoffs (κ = 0.62, 0.89, and 0.94; prevalence- and bias-adjusted κ = 0.83, 0.89, and 0.94 at 30, 50, and 100 ng/mL, respectively). Precision was acceptable (SD ≤ 6 ng/mL; coefficient of variation < 6%). Median time-to-result was shorter with the POC (14 vs 73 minutes). CONCLUSIONS: The POC system shows a high level of agreement with a laboratory anti-Xa assay and reduced turnaround time, supporting its potential use in the emergency setting.

Read the paper