A bicentric pilot study of apixaban and rivaroxaban quantification using a novel point-of-care system in the emergency department.
Cindy Tissier, Tiffany Madranges, Nicolas Peschanski, Véronique Le Meur, Thibault Donnet, Emmanuel De Maistre, Isabelle Gouin-Thibault
Emergency Department, University Hospital of Dijon, Dijon, France. · Hematology and Hemostasis Laboratory, University Hospital of Dijon, Dijon, France.
Key points
From the paper's conclusion- The POC system shows a high level of agreement with a laboratory anti-Xa assay and reduced turnaround time, supporting its potential use in the emergency setting.
Abstract
BACKGROUND: Direct oral anticoagulants (DOACs) do not require routine monitoring; however, rapid assessment of drug concentrations may be needed in critical situations. Laboratory-based anti-Xa assays are recommended but are not consistently available with turnaround times compatible with emergency care. OBJECTIVES: This study evaluated analytical performance, agreement at clinical thresholds, precision, and turnaround time of a whole-blood point-of-care (POC) system for measuring apixaban and rivaroxaban. METHODS: A preliminary laboratory study and a prospective bicentric pilot study in emergency departments were performed on DOAC-treated patients. Paired samples were analyzed using the POC system and a laboratory anti-Xa assay (STA-R Max3; Diagnostica Stago). Agreement within the analytical measuring range (23-200 ng/mL) was assessed using Passing-Bablok regression and Bland-Altman analysis. Concordance at 30, 50, and 100 ng/mL was evaluated using Cohen κ and prevalence- and bias-adjusted κ. Precision and time-to-result were also assessed. RESULTS: Excellent agreement was observed in both the preliminary laboratory study (r = 0.978; slope = 0.994; median bias = -5.5 ng/mL) and the emergency room study in the 69 analyzed patients (r = 0.965; slope = 1.013; median bias = -10.5 ng/mL). Agreement increased with higher cutoffs (κ = 0.62, 0.89, and 0.94; prevalence- and bias-adjusted κ = 0.83, 0.89, and 0.94 at 30, 50, and 100 ng/mL, respectively). Precision was acceptable (SD ≤ 6 ng/mL; coefficient of variation < 6%). Median time-to-result was shorter with the POC (14 vs 73 minutes). CONCLUSIONS: The POC system shows a high level of agreement with a laboratory anti-Xa assay and reduced turnaround time, supporting its potential use in the emergency setting.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Rapid measurement of direct oral anticoagulants for clinical decision-making.Research and practice in thrombosis and haemostasis · 2026
- Long-term analytical performance of apixaban measurements based on external quality assessment data.Thrombosis research · 2026
- Heparin-calibrated anti-factor Xa activity as a surrogate marker for direct oral anticoagulant levels in acute stroke care: a large observational single-center study.Neurological research and practice · 2026
- Real-World Assessment of Direct Oral Factor Xa Inhibitors: Dosing Appropriateness, Drug Exposure, and Inter-Platform Comparability of Chromogenic Anti-Xa Assays.Diagnostics (Basel, Switzerland) · 2026
- Development and Validation of a QuEChERS-LC-MS/MS Method for Simultaneous Determination of Four Direct Oral Anticoagulants in Citrate Plasma and Quantitative Dried Plasma Spots.Journal of separation science · 2026
- Management of bleeding and nonelective invasive procedures in patients receiving direct oral anticoagulants (DOACs): a 5-year single-center observational study.Research and practice in thrombosis and haemostasis · 2026