A targeted mass spectrometric assay for quantitative detection of fibrinogen in plasma.
Sandip Chavan, Ramesh Bokka, Shilpa Venkataraman, Anikha Bellad, Julie I Tange, Rachel R Leger, Rajiv K Pruthi, Dong Chen,
Department of Laboratory Medicine and Pathology, Mayo Clinic, Rochester, Minnesota, USA.
Key points
From the paper's conclusion- The liquid chromatography-tandem mass spectrometry-based assay provides precise, accurate quantification of fibrinogen across a wide clinical range and aligns closely with existing methods while offering analytical specificity. With further clinical validation, it could be used as a diagnostic clinical test.
Abstract
BACKGROUND: Fibrinogen is a key glycoprotein that plays a central role in clot formation, structure, and stability. Both quantitative fibrinogen deficiency and qualitative abnormalities can lead to bleeding or thrombotic disorders. Accurate quantification of fibrinogen antigen is essential for clinical evaluation; however, current immunoassays and clot-based surrogate methods are prone to interference and may lack traceability to international standards. OBJECTIVES: To develop a targeted mass spectrometry-based assay for accurate quantification of fibrinogen antigen in plasma and to compare its performance with established functional and immunoturbidimetric assays. METHODS: We developed a targeted liquid chromatography-tandem mass spectrometry-based assay using isotope-labeled peptides that accurately measures fibrinogen antigen concentration with high precision (coefficient of variation < 20%) across a broad linear dynamic range of 10.2 to 1307.5 mg/dL, enabling accurate detection of hypofibrinogenemia and hyperfibrinogenemia. Assay performance was evaluated in healthy donors (n = 47) and clinical specimens (n = 31). RESULTS: Fibrinogen antigen levels in normal donors followed a normal distribution (mean ± SD, 315 ± 42 mg/dL). The targeted assay, employing selected reaction monitoring, correlated strongly with immunoturbidimetry (Pearson's correlation coefficient [r] = .943) and the Clauss and prothrombin time-derived fibrinogen assays (r = .91 and r = .93, respectively). Stratification by clinical category confirmed elevated fibrinogen levels in thrombosis-related disorders, intermediate levels in controls, and low levels in liver disease and critical illness. CONCLUSION: The liquid chromatography-tandem mass spectrometry-based assay provides precise, accurate quantification of fibrinogen across a wide clinical range and aligns closely with existing methods while offering analytical specificity. With further clinical validation, it could be used as a diagnostic clinical test.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Congenital Afibrinogenemia and Hypofibrinogenemia: Genetic Basis, Pathophysiology, Laboratory Diagnosis, and Management.Hematology reports · 2026
- Evaluation of the Global Fibrinolytic Capacity of Whole BloodAnalytical Chemistry · 2026
- Challenges in the Quantification of Free TFPI in Human Plasma Following Marstacimab Treatment.The AAPS journal · 2026
- A-187 The A2M assay of Diagam on the DxC500AU from Beckman Coulter.Clinical Chemistry · 2026
- A-250 A MALDI mass spectrometry-based assay for the determination of ADAMTS13 activityClinical Chemistry · 2026
- Development of Highly Multiplex Targeted Proteomics Assays in Biofluids Using a Nominal Mass Ion Trap Mass SpectrometerMolecular & Cellular Proteomics · 2026