Clinical Chemistry

A-250 A MALDI mass spectrometry-based assay for the determination of ADAMTS13 activity

Olajide Olaleye, Sandip Chavan, Ramesh Bokka, Shilpa Venkataraman, Jinyong Kim, Anthony Maus, Julie Tange, Rachel Leger,

Key points

From the paper's conclusion
  • Here, we developed a highly sensitive, rapid, and reliable mass spectrometry-based assay to measure ADAMTS13 activity in plasma using matrix-assisted laser desorption/ionization (MALDI) mass spectrometry.

Abstract

Background Severe deficiency of ADAMTS13 (A Disintegrin and Metalloproteinase with Thrombospondin motif type 1 member 13) enzyme activity can lead to thrombotic thrombocytopenic purpura, a life-threatening disorder. Prompt diagnosis and initiation of plasma exchange is necessitated upon clinical presentation. ADAMTS13 activity measurement relies on detecting the cleavage of von Willebrand Factor (VWF) at the A2 domain by the enzyme in test plasma. Most clinical tests use a 73-amino acid peptide segment of VWF (VWF73) as substrate and are fluorescence-based. However, their results sometimes do not concur with the clinical presentation. Here, we developed a highly sensitive, rapid, and reliable mass spectrometry-based assay to measure ADAMTS13 activity in plasma using matrix-assisted laser desorption/ionization (MALDI) mass spectrometry. Methods Test plasma, calibrators, or controls were incubated with VWF73 substrate for 30 min in Bis-Tris buffer, together with a stable-isotope labeled decapeptide as internal standard. Calibrators, along with controls, were prepared from a commercial pooled normal human plasma and standardized against the WHO International Standard ADAMTS13 Plasma 12/252 and were used to construct an eight-point calibration curve. SPE clean-up of reaction mixture was achieved under a positive nitrogen pressure of about 5 mmHg using Strata-X 33 µm polymeric reversed phase, 10 mg sorbent per well. The eluate was spotted onto the target plate, topped by a layer of the matrix a-cyano-4-hydroxycinnamic acid. Spectral data was acquired in positive ionization mode on a rapifleX MALDI Tissuetyper (Bruker). For quantification, intensity of target decapeptide was normalized by intensity of internal standard. The method was evaluated using 191 plasma samples, including 121 patient samples, along with a World Health Organization (WHO) international standard. Results Target decapeptide was generated upon incubation of VWF73 with plasma, which was measured by MALDI mass spectrometry, along with the internal standard. Total assay time was nearly one hour, with an assay range of 2–109 IU/dL. Intra-assay coefficient of variation (CV) was <2% and inter-assay CV was <6%. The throughput matched the current fluorescence resonance energy transfer (FRET) assay workflow. Based on the 10 IU/dL threshold for TTP, the results showed high agreement (Kappa coefficient, 0.89) with clinical testing using FRETS-VWF73 assay. Conclusion Our MALDI-MS assay combines rapid and minimal sample preparation with high sensitivity and reproducibility that has potential to improve patient care.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers