Mechanical ventilation vs bag-valve ventilation during cardiopulmonary resuscitation: a randomized crossover noninferiority trial.
S Vinay, S Manu Ayyan, V T Amrithanand, B G Kowsthubha
Department of Emergency Medicine and Trauma, Jawaharlal Institute of Postgraduate Medical Education and Research (JIPMER), Dhanvantari Nagar, Puducherry 605006, India.
Key points
From the paper's conclusion- Protocolized MV produced lower and less variable device-calculated VTi than BVV but did not meet noninferiority. BVV had a wider upper tail, whereas MV more often produced values below 6 mL/kg. The unvalidated analyzer and absence of clinical outcomes limit interpretation.
Abstract
AIM: Evidence guiding mechanical ventilation (MV) during CPR is limited, and device-calculated inspired tidal volume (VTi) during ongoing chest compressions may differ from the set volume during compressions. We compared VTi with protocolized MV and bag-valve ventilation (BVV) through an endotracheal tube for noninferiority. METHODS: This single-center randomized crossover noninferiority trial enrolled 104 intubated adults meeting termination-of-resuscitation criteria without return of spontaneous circulation to receive BVV and MV in randomized order within a 5-min peri-termination window. Device-calculated VTi was recorded for 1 min per method during ongoing compressions. The primary outcome was the between-treatment difference in per-participant median VTi; the noninferiority margin was -150 mL (MV minus BVV). RESULTS: Among 99 participants completing both periods, mean per-participant median VTi was 514.4 mL with BVV and 368.0 mL with MV; coefficients of variation were 40.1% and 28.3%. The adjusted mean difference was -146.2 mL (95% CI, -187.0 to -105.5). Because the lower bound crossed -150 mL, noninferiority was not demonstrated. Median VTi exceeded 10 mL/kg predicted body weight in 31.3% with BVV and 0% with MV, whereas values below 6 mL/kg occurred in 26.3% and 45.5%. CONCLUSIONS: Protocolized MV produced lower and less variable device-calculated VTi than BVV but did not meet noninferiority. BVV had a wider upper tail, whereas MV more often produced values below 6 mL/kg. The unvalidated analyzer and absence of clinical outcomes limit interpretation. CLINICAL TRIAL REGISTRATION: Clinical Trials Registry - India; CTRI/2024/08/072738.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Mechanical versus manual bag-valve ventilation during cardiopulmonary resuscitation in adults with cardiac arrest: A systematic review and Bayesian meta-analysis.Resuscitation · 2026
- Comparison of bag-valve-device ventilation and volume-controlled ventilation with laryngeal tube evo in intra-arrest ventilation-a prospective randomized study in human body donors.Journal of anesthesia, analgesia and critical care · 2026
- Accuracy of ventilation feedback devices compared with a reference standard during asynchronous ventilation with continuous chest compressions in a simulated cardiac arrest: a prospective non-randomized cross-over study.Resuscitation plus · 2026
- Positive end-expiratory pressure versus sham valve/zero end-expiratory pressure in cardiopulmonary resuscitation during manual ventilation to improve neurological outcomes in adult patients suffering an out-of-hospital cardiac arrest - an investigator-initiated, pragmatic, registry-based, multicenter, parallel-group, triple-blind randomized controlled superiority clinical trial in the ARREST registry (REVIVE-PEEP protocol Stage-1 Registered Report)medRxiv · 2026
- Reversed airflow during asynchronous volume controlled mechanical ventilation in Out-of-Hospital Cardiac Arrest: a physiological observational study.Resuscitation · 2026
- Real-time feedback during bag-valve-mask ventilation for out-of-hospital cardiac arrest: a before-and-after study (VEntilation during Cardiac Arrest Resuscitation Study, VECARS-2).Resuscitation · 2026