Visualizing ventriculoperitoneal shunt valve settings with computed tomography, using extended CT scale reconstructions.
E Löfblad, P E Åslund, S Hagren, M Segelsjö, D Fällmar
Uppsala University, Department of Surgical Sciences, Neuroradiology, Uppsala, Sweden. Electronic address: erik.lofblad@akademiska.se. · Uppsala University Hospital, Department of Medical Physics, Uppsala, Sweden.
Key points
From the paper's conclusion- ECTS reconstructions with a small FOV, derived from routine clinical CT head examinations, allowed reliable readout of programmable shunt valve settings for Codman Hakim, Medtronic Strata, and proGAV 2.0 valves. These findings represent an important step toward a unified CT-based approach for ventriculoperitoneal shunt evaluation.
Abstract
INTRODUCTION: Ventriculoperitoneal shunts are commonly evaluated using radiographs to confirm valve settings and catheter integrity. Although modern CT provides detailed soft tissue assessment, valve setting readouts have been unreliable due to insufficient spatial resolution and attenuation range. This study investigated whether image reconstructions with extended CT scale (ECTS) can enable reliable identification of valve settings for three common programmable shunt valve types on routine clinical CT examinations. METHODS: Fifteen routine head CT examinations from 14 patients with programmable shunt valves (five each of Codman Hakim, Medtronic Strata and Miethke proGAV 2.0) were included. For each examination, multiple reconstructions were generated from raw data, including standard bone-kernel images and ECTS-based reconstructions. One representative image from each reconstruction was evaluated by 15 blinded reviewers, who recorded valve setting and confidence. Accuracy was assessed relative to clinically documented reference settings, with ±1 step considered an acceptable deviation. RESULTS: All valve settings were correctly identified within ±1 step on ECTS reconstructions across all three valve types. ECTS significantly improved setting readout compared with standard reconstructions. Neither iterative reconstruction nor metal-artifact reduction provided additional benefit. CONCLUSION: ECTS reconstructions with a small FOV, derived from routine clinical CT head examinations, allowed reliable readout of programmable shunt valve settings for Codman Hakim, Medtronic Strata, and proGAV 2.0 valves. These findings represent an important step toward a unified CT-based approach for ventriculoperitoneal shunt evaluation. IMPLICATIONS FOR PRACTICE: Head CT is commonly performed in patients with programmable shunts, often to rule out unintended valve settings or dysfunction. Every increased possibility to read the setting directly on CT, such as adding ECTS reconstructions, will directly benefit the patients through increased safety and decreased or waived need for additional projection radiography.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Clinical utility of SPECT/CT radionuclide shuntgraphy in suspected ventriculoperitoneal shunt dysfunction.Acta neurochirurgica · 2026
- EIT reconstruction parameters’ impact on ventilation monitoring featuresCurrent Directions in Biomedical Engineering · 2026
- Programmable shunt valve stability after MRI: a 2-year institutional experience across valve types and field strengths.Child's nervous system : ChNS : official journal of the International Society for Pediatric Neurosurgery · 2026
- Intraoperative cone-beam CT for real-time quality control in CSF shunt surgery: a prospective workflow study.Brain & spine · 2026
- Longitudinal boundary-anchored CT profiling of the frontal horns after cerebrospinal fluid shunting in idiopathic intracranial hypertension: a pilot study.World neurosurgery · 2026
- Feasibility of a 2-Minute Multi-Echo UTE Acquisition for Simultaneous CT-Like Bone-Weighted Imaging and Quantitative T2* Mapping of Short-T2 TissuemedRxiv · 2026