Biomechanical evaluation of shield design in multiple anterior implant restorations: A three-dimensional finite element analysis.
Qianqian Zuo, Haidong Teng, Yuanli Zhang, Zhan Liu
Graduate student, Department of Biomedical Engineering, School of Architecture and Environment, Sichuan University, Chengdu, PR China; and Graduate student, Sichuan University Yibin Park, Yibin, PR China. · Graduate student, Department of Mechanics, School of Architecture and Environment, Sichuan University, Chengdu, PR China; and Graduate student, Sichuan University Yibin Park, Yibin, PR China.
Key points
From the paper's conclusion- The Half-C shield provided biomechanical advantages by reducing peak stress and moderating strain distribution, indicating better mechanical performance than the classic shield and supporting its use in 2-tooth restorations.
Abstract
STATEMENT OF PROBLEM: The socket shield technique (SST) has demonstrated benefits in preserving esthetic outcomes. However, the biomechanical behavior associated with its application in multiple adjacent implant restorations remains unclear. PURPOSE: The purpose of this finite element analysis study was to evaluate the biomechanical influence of classic and Half-C shield designs in bilateral maxillary central incisor and combined central-lateral incisor implant restorations and to provide biomechanical evidence to support shield selection in multitooth SST applications. MATERIAL AND METHODS: A 3-dimensional model of the complete maxilla was reconstructed. Two common anterior implant restoration scenarios were simulated: bilateral central incisors and combined central-lateral incisors. Classic and Half-C shields were generated, and 8 finite element models were constructed. Time-dependent loading was applied to the crowns. Stress distributions in the implant region and strain distributions in the surrounding bone were compared. Statistical analysis was performed using a general linear model univariate procedure (α=.05). RESULTS: The Half-C shield exhibited lower overall stress and produced a more balanced strain distribution than the classic shield. Shield type and socket aspects both significantly influenced the strain in the surrounding bone (P<.001). The Half-C configuration reduced mesiodistal strain and showed synergistic stress-relieving effects when applied on both teeth. CONCLUSIONS: The Half-C shield provided biomechanical advantages by reducing peak stress and moderating strain distribution, indicating better mechanical performance than the classic shield and supporting its use in 2-tooth restorations.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Finite element evaluation of the biomechanical feasibility of a multi-shield socket shield technique for a vertically fractured maxillary central incisor.Medical engineering & physics · 2026
- Influence of all-on-4 implant angles on stability and crestal bone - A finite element optimization analysis.Journal of dental sciences · 2026
- Biomechanical Evaluation of Different Posterior Mandibular Implant Configurations Under Bruxism: An FEA Study.Journal of functional biomaterials · 2026
- Single Versus Double Implants in a Narrowed Mandibular Molar Space: A Finite Element Analysis.Journal of functional biomaterials · 2026
- Biomechanical evaluation of surface texturing and hybrid coatings in dental implants: A finite element analysis of stress distribution under bone loss conditions.Proceedings of the Institution of Mechanical Engineers. Part H, Journal of engineering in medicine · 2026
- Biomechanical Interaction Between Endosseous and Patient-Specific Subperiosteal Implants in Full-Arch Rehabilitation of the Atrophic Maxilla: A Finite Element Analysis.The International journal of oral & maxillofacial implants · 2026