Influence of gating and drift correction on prostate cancer treatment margins on a 1.5 T magnetic resonance linear accelerator.

Nienke D Sijtsema, Guus Grimbergen, Eline N de Groot-van Breugel, Jochem R N van der Voort van Zyp, Johannes C J de Boer

Department of Radiation Oncology, Division of Imaging & Oncology, University Medical Center Utrecht, Utrecht, the Netherlands.

Key points

From the paper's conclusion
  • In this study, intra-fraction motion management using gating and baseline shift corrections on a 1.5 T magnetic resonance linear accelerator was evaluated in 93 prostate cancer patients.

Abstract

In this study, intra-fraction motion management using gating and baseline shift corrections on a 1.5 T magnetic resonance linear accelerator was evaluated in 93 prostate cancer patients. Motion tracks were analyzed to assess the frequency, size and duration of baseline shift corrections and derive planning target volume (PTV) margins. Baseline shift corrections were required in 89.2% of patients, with a range of 0-6 corrections per fraction. The resulting PTV margins were 1.2 mm left-right, and 2.3 mm anterior-posterior and cranial-caudal. Because these values were conservatively estimated, a 2 mm isotropic PTV margin was clinically implemented without increasing treatment time.