Dosimetric Impact of Residual Intrafraction Motion in Gated MR-Guided Prostate SBRT With Focal Dose Intensification
K. H. Tam, S. Jogi, A. Sobremonte, J. D. Ohrt, D. J. Rhee, J. Yang, Y. Ding, K. K. Brock,
Key points
From the paper's conclusion- Purpose: To quantify the dosimetric effect of residual intrafraction motion during gated MR-guided prostate stereotactic body radiotherapy with simultaneous integrated boost to the dominant intraprostatic lesion, and to derive and evaluate target-specific, population-based margins designed to mitigate motion-induced coverage loss.
Abstract
Purpose: To quantify the dosimetric effect of residual intrafraction motion during gated MR-guided prostate stereotactic body radiotherapy with simultaneous integrated boost to the dominant intraprostatic lesion, and to derive and evaluate target-specific, population-based margins designed to mitigate motion-induced coverage loss. Materials and Methods: Thirty patients were treated on a 1.5-T MR-Linac with 36.25, 40, and 45 Gy prescribed to the planning target volume, clinical target volume (CTV), and gross tumor volume (GTV), in 5 fractions. Across 150 fractions, segment-level dose matrices synchronized with time-resolved target displacements were combined to reconstruct per-fraction and cumulative dose. Directional prescription-isodose deviations were used to derive internal target volume margins, which were evaluated through offline replanning, gating-envelope analysis, and dose reconstruction. Results: Median 95th-percentile motion ranges were 0.5, 1.4, and 2.9 mm in the left-right, anterior-posterior, and superior-inferior directions; these exceeded 3 mm in 0.7%, 13.2%, and 47.7% of fractions, respectively. The median treatment time was 13.2 min (IQR, 12.0-15.1 min) with a median gating duty cycle of 93.2% (80.6%-97.8%). Per-fraction median GTV deviations were -1.2% (-2.4% to -0.1%) for D95 and -5.6% (-15.3% to -0.2%) for V45Gy; corresponding CTV deviations were -0.5% (-1.1% to -0.1%) and -2.1% (-4.1% to -0.4%). Cohort-derived margins of 2/2/1/2/1/2 mm for the GTV and 1/1/1/2/1/2 mm for the CTV in the left/right/anterior/posterior/superior/inferior directions achieved sufficient coverage in approximately 94% of fractions in each direction. Gated delivery using the proposed margins improved coverage but may reduce duty cycle by an average of 6.9 +/- 4.7 %. Conclusion: Residual intrafraction motion produced directionally asymmetric coverage loss, with the largest effect on the boosted GTV. Motion-inclusive dose reconstruction enabled derivation of practical asymmetric margins that improved robustness while identifying the tradeoff between target coverage and delivery efficiency.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Influence of gating and drift correction on prostate cancer treatment margins on a 1.5 T magnetic resonance linear accelerator.Physics and imaging in radiation oncology · 2026
- Optimal gating envelope size for automatic gating in magnetic resonance imaging-guided radiotherapy.Physica medica : PM : an international journal devoted to the applications of physics to medicine and biology : official journal of the Italian Association of Biomedical Physics (AIFB) · 2026
- A real-time image-guided six-degree-of-freedom internal-external correlation framework for motion monitoring on a standard linear accelerator.Physics in medicine and biology · 2026
- Dosimetric impact of baseline shift plan in a magnetic resonance linear accelerator according to target motion and plan characteristicsJournal of Applied Clinical Medical Physics · 2026
- Dosimetric Evaluation of MR-Guided IMRT and VMAT Techniques for Early-Stage Glottic Cancer: A Pilot Comparative Planning Study.Medicina (Kaunas, Lithuania) · 2026
- A novel automatic real-time motion tracking method in MRI-guided radiotherapy using enhanced Tracking-Learning-Detection framework with automatic segmentation.Radiation oncology (London, England) · 2026