Target-volume-based prescription selection for specified adverse-effect levels in automated stereotactic radiosurgery.
Joel A Pogue, Rex Cardan, Farnaz Rahim Li, Rodney Sullivan, Natalie Viscariello, Luke Moradi, John Fiveash, Richard Popple
Department of Radiation Oncology, The University of Alabama at Birmingham, 619 19th Street South, Birmingham, AL 35249, USA.
Key points
From the paper's conclusion- We derived a direct isoeffect prescription dose ( D RX ) model from previously validated target-volume ( V target ) power laws for single-isocenter automated SRS.
Abstract
Automated stereotactic radiosurgery (SRS) planning produced consistent dose falloff, enabling adverse-effect-relevant isodose volumes ( IDV ) to be estimated before plan generation. We derived a direct isoeffect prescription dose ( D RX ) model from previously validated target-volume ( V target ) power laws for single-isocenter automated SRS. IDV coefficients and exponents were modelled as continuous functions of relative isodose level ( RIL ), allowing D RX to be solved from V target , desired IDV , and HyTEC brain adverse-effect dose thresholds. Across held-out levels, target-level IDV residuals had a median of 0.2%, an interquartile range of [-5.2%, +7.0%], and a 95% interval of [-12.8%, +27.0%].
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Benchmarking intermediate dose spillage in single target stereotactic radiosurgery for brain metastases.Physics and imaging in radiation oncology · 2026
- In silico assessment of out-of-field dose in multiple brain metastases stereotactic radiosurgery.Physical and engineering sciences in medicine · 2026
- Evaluation of EPID-Based Transmission and DLG Correction Methods for Dosimetric Verification in HyperArc Single-Isocenter Multiple Target Radiosurgery.Diagnostics (Basel, Switzerland) · 2026
- The Role of Biologically Effective Dose in Clinical Improvement and Hyper-Response After Gamma Knife Radiosurgery for Essential Tremor.Neurosurgery · 2026
- Integrated isocenter and sector duration optimization (ISDO) for gamma knife radiosurgery.Physics in medicine and biology · 2026
- Dose deposition characteristics in theoretical 4π and clinical limiting 2π geometry: A proof of concept and potential implications for dose escalation.Journal of cancer research and therapeutics · 2026