Quantifying manual editing time in autocontouring via audit log analysis.
Geert De Kerf, Gabriele Balletti, Nina Menten, Michaël Claessens, Thibaut D'homme, Joachim Marichal, Mark J Gooding, Dirk Verellen
Department of Radiation Oncology, Iridium Netwerk, Wilrijk (Antwerp), Belgium. · Machine-Learning Department, RaySearch Laboratories, Stockholm, Sweden.
Key points
From the paper's conclusion- Vendor audit logs could be translated into accurate structure-level editing time estimates, enabling low-burden monitoring of manual workload. Editing strategy influenced time-geometry relationships, supporting log-based time endpoints alongside geometric metrics for meaningful auto-contouring evaluation.
Abstract
BACKGROUND AND PURPOSE: Artificial intelligence (AI)-based contouring reduces delineation workload, yet manual corrections remain required. Reliable, scalable quantification of structure-level editing time is needed to evaluate clinical usability beyond geometric similarity metrics as geometric metrics may not reflect how tooling impacts edit time. MATERIALS AND METHODS: This retrospective observational study analysed 3083 AI-generated, edited structures created in routine practice. Structure-modifying actions were extracted from the treatment planning system audit log database and converted to per-structure editing time by summing inter-event intervals. Automatic times were benchmarked against manual time recordings for 54 structures across 8 patient cases using Lin's concordance correlation coefficient (CCC). Added path length (APL) between AI and corrected contours was computed, and the associations between editing time, APL and tooling (interpolation use) were investigated using Spearman's correlation coefficient. RESULTS: Audit log-derived editing time agreed well with manual timing (CCC = 0.93). In the full cohort, 3083/24970 structures were edited (12% correction rate). Brush tools were used in 80% and contour interpolation in 42% of edited structures. Median editing time and APL were higher with interpolation than without (81 s vs 58 s; 232 mm vs 18 mm; both p < 0.05), and correlation between editing time and APL was fair (r = 0.38 vs 0.51). CONCLUSIONS: Vendor audit logs could be translated into accurate structure-level editing time estimates, enabling low-burden monitoring of manual workload. Editing strategy influenced time-geometry relationships, supporting log-based time endpoints alongside geometric metrics for meaningful auto-contouring evaluation.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Teaching organ-at-risk delineation in the AI era: educational impact and dosimetric relevance of a structured training interventionBMC Medical Education · 2026
- Clinical validation of AI‐assisted contouring in prostate radiation therapy treatment planning: Highlighting automation bias and the need for standardized quality assuranceJournal of Applied Clinical Medical Physics · 2025
- A prospective randomized evaluation of AI assisted GTV contouring in head-and-neck radiotherapy.Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology · 2026
- Organ-at-risk contouring education in the era of AI-assisted practice: Insights from an Australian undergraduate radiation therapy program.Journal of medical imaging and radiation sciences · 2026
- Performances of AI-based automatic delineation of CTV and Organs-at-Risk for breast radiotherapy planning: including inter-observer variability in the evaluation process.Physica medica : PM : an international journal devoted to the applications of physics to medicine and biology : official journal of the Italian Association of Biomedical Physics (AIFB) · 2026
- ROECSG report: Real‐world implementation patterns and perceptions of artificial intelligence in radiation oncologyJournal of Applied Clinical Medical Physics · 2026