Robust spectral averaging and chemical shift calculation for hyperpolarized 129Xe MR spectroscopy.
Seth N Lee, Suphachart Leewiwatwong, Anna Costelle, David Mummy, Bastiaan Driehuys
Medical Physics Graduate Program, Duke University, 2424 Erwin Rd Suite 101, Durham, NC 27705, USA. · Department of Biomedical Engineering, Duke University, 1427 Fitzpatrick Center, Box 90281, Durham, NC 27708, USA.
Key points
From the paper's conclusion- RBC:M is most repeatable when averaged over a cardiac cycle, although 1-s averaging yields similar performance. Dissolved-phase shifts are most repeatable using dedicated gas-phase excitations as the 0-ppm reference frequency.
Abstract
PURPOSE: To establish methods for 129Xe MRS quantification of the red blood cell to membrane signal ratio (RBC:M) and dissolved-phase chemical shifts that are optimized for repeatability. METHODS: The effects of cardiogenic RBC signal oscillations on RBC:M were simulated to determine the optimal FID averaging duration; variability was evaluated in healthy volunteers and idiopathic pulmonary fibrosis patients (n = 55). Dissolved-phase shifts were calculated in progressive pulmonary fibrosis patients (n = 114) using two 0-ppm references: the incidental gas resonance arising during dissolved-phase excitation and a dedicated gas-phase excitation. The repeatability of each method was compared, and that of the dedicated gas reference was assessed in subjects with various cardiopulmonary disorders (n = 155). Optimized methods were then used to re-evaluate published healthy reference values. RESULTS: The variability of RBC:M measurements followed a sinc-like envelope with FID averaging duration, exhibiting minima at integer multiples of the cardiac period. Dissolved-phase shifts were most repeatable when referenced to the dedicated gas signal. Cardiac cycle averaging did not change the healthy reference RBC:M (0.49 ± 0.11, p = 0.75), but using the dedicated gas resonance increased healthy reference membrane and RBC shifts by 0.2 and 0.1 ppm to 197.8 ± 0.3 and 218.3 ± 0.5 ppm, respectively (p < 0.001). Membrane and RBC shifts demonstrated good repeatability coefficients of 0.32 and 0.52 ppm. CONCLUSION: RBC:M is most repeatable when averaged over a cardiac cycle, although 1-s averaging yields similar performance. Dissolved-phase shifts are most repeatable using dedicated gas-phase excitations as the 0-ppm reference frequency.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Correlating Pulmonary Hemodynamics With HP 129Xe Dissolved-Phase MRSI in a Controlled Porcine Intervention Study.Magnetic resonance in medicine · 2026
- Advancing oxygen-enhanced MRI acquisition and modelling through dynamic ${\Delta}R_{1}$ and ${\Delta}R_2^*$ sensingarXiv · 2026
- Feasibility of Singular Value Decomposition Implementation for 129Xe Hyperpolarized Chemical Exchange Saturation Transfer (HyperCEST) Molecular Imaging.Magnetic resonance in medicine · 2026
- Time-Resolved Monitoring of Surface-Adsorbed H2O Migration into Mesoporous Media at Submillimeter Spatial Resolution through Hyperpolarized Xenon-129 NMR SpectroscopyThe Journal of Physical Chemistry C · 2026
- Evaluations of retrospective frequency and phase correction methods for single-voxel MR spectroscopy at 7T.bioRxiv : the preprint server for biology · 2026
- Hyperpolarized Xenon 129 MRI to Evaluate Regional Gas-Exchange Abnormalities in Cystic Fibrosis Lung Disease.Radiology. Cardiothoracic imaging · 2026