Cost-effective 0.5 T benchtop NMR/MRI system with spatial inhomogeneity compensation and digital field-frequency lock.
Kaisheng Lin, Miaotian Wang, Shiwei Qin, Weimin Wang, Kai Zhong, Wenhui Yang, Qiuliang Wang
Institute of Electrical Engineering, Chinese Academy of Sciences, Beijing, 100190, China; School of Electronic, Electrical and Communication Engineering, University of Chinese Academy of Sciences, Beijing, 100190, China; Beijing Key Laboratory of Applied Superconductivity, Institute of Electrical Engineering, Chinese Academy of Sciences, Beijing, 100190, China. Electronic address: linkaisheng@mail.iee.ac.cn. · School of Electronics, Peking University, Beijing, 100190, China. Electronic address: wmt@pku.edu.cn.
Key points
From the paper's conclusion- These results demonstrate an integrated approach for mitigating both spatial field inhomogeneity and temporal field instability in compact permanent-magnet NMR/MRI systems.
Abstract
Low-field benchtop magnetic resonance systems offer advantages in cost, footprint, and accessibility, but their performance is often constrained by spatial magnetic-field inhomogeneity, temporal field drift, and limited sequence-control flexibility. This work presents a highly integrated 0.5 T benchtop NMR/MRI platform based on the open-source MaRCoS framework and a customized PyPulseq-compatible control workflow. The system incorporates programmable RF and gradient control, a 12-channel high-order active shim array, a dual-nuclear 1H/19F RF probe, and 19F-referenced digital field-frequency locking (FFL). Active shimming reduced the spectral linewidth to 0.35 ppm. To further address broadening caused by residual spatial B0 variations, imaged reference deconvolution was applied to anhydrous-ethanol spectroscopy, recovering fine J-coupled structures that were unresolved by direct spatial accumulation. Temporal field instability was addressed using digital FFL, in which 19F frequency offsets were used to update the 1H carrier frequency; long-term FID experiments demonstrated improved spectral stability and coherent averaging under deliberately imposed field drift. The platform also supported two-dimensional spin-echo MRI, with a nominal in-plane pixel size down to 0.059 × 0.076 mm2. During extended spin-echo acquisitions, digital FFL reduced drift-induced displacement and blurring. These results demonstrate an integrated approach for mitigating both spatial field inhomogeneity and temporal field instability in compact permanent-magnet NMR/MRI systems.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- A closer look at magnetic field corrections for a 0.35 T MRI‐LinacMedical Physics · 2026
- A Reference System for Open Source Portable Low-Field MRIarXiv · 2026
- Free-Breathing PDFF and R2* Quantification at 0.55T: Reduced Respiratory Motion Sensitivity and a Time-Resolved 4D MRI ApproacharXiv · 2026
- From Repeatability to Reproducibility in Quantitative Low-Field MRI: A Metrology-Informed Framework for Phantom Evidence.Journal of imaging · 2026
- Diffusion broadening of the point spread function in steady-state MRIarXiv · 2026
- OpenMRF: A Modular, Vendor-Neutral Open-Source Framework for Magnetic Resonance Fingerprinting Using Pulseq.Magnetic resonance in medicine · 2026