A closer look at magnetic field corrections for a 0.35 T MRI‐Linac
Barnali Brahma, Austen N. Curcuru, Taeho Kim, Deshan Yang, H Michael Gach
Key points
From the paper's conclusion- Therefore, FICO is beneficial for minimizing off‐resonances.
Abstract
Background Magnetic FIeld COrrection (FICO) was implemented on a 0.35 T MRI‐Linac to minimize B 0 off‐resonances. FICO parameters were derived from phantom measurements conducted at 30° gantry angle increments. FICO data were interpolated for every gantry angle, stored in a look‐up table, and read during MRI‐Linac simulation and treatment. Purpose Verify the performance and efficacy of FICO. Methods We measured off‐resonances in a 24 cm diameter spherical volume phantom and compared the FICO values to our measurements conducted at every: (1) gantry angle for center frequency and gradient shim; (2) 2° increment for B 0 eddy currents; and (3) 10° increment for gradient and cross‐term eddy currents. FICO results were compared with pre‐FICO methods using the settings from a single home gantry angle (HGA), for example, 300°. Results Off‐resonances decreased from a mean of ‐20.19 (HGA) to ‐0.61 Hz (FICO). Dispersion decreased from 81.10 Hz (HGA) to 52.71 Hz (FICO) and 28.0 Hz (FICO with gradient shim updates). B 0 eddy currents remained within specification (± 0.1 µT∙m/mT) with FICO. X‐ B 0 and Y‐ B 0 eddy currents exceeded the limit when using HGA. Gradient and cross‐term eddy currents were within limits (± 0.1%) for both FICO and HGA. X and cross‐term (Y→X, and X→Z) eddy currents were significantly smaller with FICO than with HGA ( p ≤ 0.02). Gradient eddy currents generally dropped by 24% using FICO versus HGA. X and Y gradient shim root mean square errors dropped by half by reducing the gantry angle increment from 30° to 20°. Conclusions FICO resulted in reductions in center frequency offsets and full width at half maximum (FWHM) values, gradient shim amplitudes, and B 0 eddy currents. Therefore, FICO is beneficial for minimizing off‐resonances. However, FICO will benefit from using smaller gantry angle increments for interpolations and updating the gradient shim amplitudes and center frequency offsets before each treatment gantry angle.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Characterizing magnetic field coupling in MRI-guided radiotherapy by solving a Dirichlet problem with sparsely sampled boundary conditions.Physics in medicine and biology · 2026
- Open-Source Framework for Measurement of the Gradient Impulse Response FunctionbioRxiv · 2026
- Frame-wise multi-echo distortion correction for superior functional MRIImaging Neuroscience · 2026
- Cost-effective 0.5 T benchtop NMR/MRI system with spatial inhomogeneity compensation and digital field-frequency lock.Journal of magnetic resonance (San Diego, Calif. : 1997) · 2026
- ReCoFlow-MRI: Few-Step Conditional Flow Matching for GNL-Induced Nonlinear MRI ReconstructionarXiv · 2026
- Fat Fraction Mapping in the Presence of Magnetic Field Inhomogeneities With Phase‐Cycled bSSFP at 3 TNMR in Biomedicine · 2026