A closer look at magnetic field corrections for a 0.35 T MRI‐Linac

Barnali Brahma, Austen N. Curcuru, Taeho Kim, Deshan Yang, H Michael Gach

Key points

From the paper's conclusion
  • Therefore, FICO is beneficial for minimizing off‐resonances.

Abstract

Background Magnetic FIeld COrrection (FICO) was implemented on a 0.35 T MRI‐Linac to minimize B 0 off‐resonances. FICO parameters were derived from phantom measurements conducted at 30° gantry angle increments. FICO data were interpolated for every gantry angle, stored in a look‐up table, and read during MRI‐Linac simulation and treatment. Purpose Verify the performance and efficacy of FICO. Methods We measured off‐resonances in a 24 cm diameter spherical volume phantom and compared the FICO values to our measurements conducted at every: (1) gantry angle for center frequency and gradient shim; (2) 2° increment for B 0 eddy currents; and (3) 10° increment for gradient and cross‐term eddy currents. FICO results were compared with pre‐FICO methods using the settings from a single home gantry angle (HGA), for example, 300°. Results Off‐resonances decreased from a mean of ‐20.19 (HGA) to ‐0.61 Hz (FICO). Dispersion decreased from 81.10 Hz (HGA) to 52.71 Hz (FICO) and 28.0 Hz (FICO with gradient shim updates). B 0 eddy currents remained within specification (± 0.1 µT∙m/mT) with FICO. X‐ B 0 and Y‐ B 0 eddy currents exceeded the limit when using HGA. Gradient and cross‐term eddy currents were within limits (± 0.1%) for both FICO and HGA. X and cross‐term (Y→X, and X→Z) eddy currents were significantly smaller with FICO than with HGA ( p ≤ 0.02). Gradient eddy currents generally dropped by 24% using FICO versus HGA. X and Y gradient shim root mean square errors dropped by half by reducing the gantry angle increment from 30° to 20°. Conclusions FICO resulted in reductions in center frequency offsets and full width at half maximum (FWHM) values, gradient shim amplitudes, and B 0 eddy currents. Therefore, FICO is beneficial for minimizing off‐resonances. However, FICO will benefit from using smaller gantry angle increments for interpolations and updating the gradient shim amplitudes and center frequency offsets before each treatment gantry angle.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers

A closer look at magnetic field corrections for a 0.35 T MRI‐Linac — Science Skim