From Repeatability to Reproducibility in Quantitative Low-Field MRI: A Metrology-Informed Framework for Phantom Evidence.
Helena Kolarova, Jitka Hanusova, Jan Kavalirek, Martin Halaj, Jiri Tesar
Czech Metrology Institute, Okružní 31, 638 00 Brno, Czech Republic. · KAR BioTech s.r.o., Náměstí Sítná 3105, 272 01 Kladno, Czech Republic.
Key points
From the paper's conclusion- We conducted a targeted synthesis of empirical low-field reports, quantitative-imaging methods, metrology guidance, and MRI standards using a documented search and explicit source-classification criteria.
Abstract
Low-field magnetic resonance imaging includes portable and open architectures, but quantitative results remain sensitive to field strength, temperature, environment, system state, acquisition, reconstruction, and model fitting. This methodological review asks what evidence supports claims of repeatability, intermediate precision, reproducibility, qualified reference-linked bias, measurement uncertainty, and use relevance. We conducted a targeted synthesis of empirical low-field reports, quantitative-imaging methods, metrology guidance, and MRI standards using a documented search and explicit source-classification criteria. The resulting framework separates three independent axes: represented precision conditions, strength of reference linkage, and relevance to the intended technical or clinical use. Each claim is linked to an experimental design, statistical estimand, configuration record, uncertainty model, and decision rule. An explicitly synthetic low-field T1 example illustrates design identifiability, precision-based replication planning, mixed-effects analysis, an uncertainty budget, and a guard-banded technical decision; it is not experimental validation. A staged pathway allows laboratories to begin with provenance, constancy, and within-system repeatability before progressing to reference-linked or multicentre studies. Meeting a phantom threshold establishes only the specified technical property under the tested conditions; it does not establish diagnostic utility, clinical benefit, regulatory conformity, or CE marking. The framework is an operational synthesis, not a replacement standard or validated quality score.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- A Reference System for Open Source Portable Low-Field MRIarXiv · 2026
- Integrating Regulatory Planning and Metrology into Open-Source Medical Device Research: A Proposed Practice-Informed Framework from the A4IM Low-Field MRI Project.Bioengineering (Basel, Switzerland) · 2026
- Towards the experimental assessment of exposure to extremely low frequency magnetic field.Physics in medicine and biology · 2026
- Phantom studies in medical imaging (PSMI): a guide with recommendations and checklistEuropean Radiology Experimental · 2025
- MRpro: Open Framework for Model‐Based, Learned, and Quantitative MR Imaging—Application to Low‐Field MRINMR in Biomedicine · 2026
- Repeatable tract-based diffusion metrics on a portable 64 mT MRI: a foundation for global population neuroscience.Frontiers in neuroimaging · 2026