Journal of magnetic resonance (San Diego, Calif. : 1997)

Frequency Selective Surface (FSS) radiofrequency shield for low-field MRI.

Jun-Qi Yang, Meena Rajendran, Minxuan Xu, Xudong Chen, Shao Ying Huang, Wenwei Yu

Center for Frontier Medical Engineering, Chiba University, Inage Ku, Yayoi Cho, 1-33, Chiba, 263-8522, Japan. · Singapore University of Technology and Design, 8 Somapah Road, Singapore 487372, Singapore.

Key points

From the paper's conclusion
  • The results show that the FSS suppresses coupled external interference by approximately 20dB while maintaining substantially higher B1 sensitivity than the copper shield, yielding the highest image SNR among the three configurations.

Abstract

In this paper, we propose an inductive Frequency Selective Surface (FSS) shield for RF transceiver coils in portable low-field MRI, aiming to balance electromagnetic interference suppression and coil efficiency. A high-pass mesh cylinder is designed for a 4.3MHz solenoid in a 100mT portable scanner. Shield performance is quantified in terms of shielding effectiveness (SE) and B1 sensitivity using analytical calculations and full-wave CST simulations. The study is further validated through bench measurements and MRI scans of a cylindrical phantom and an apple, with unshielded and copper-shielded solenoids serving as references. The results show that the FSS suppresses coupled external interference by approximately 20dB while maintaining substantially higher B1 sensitivity than the copper shield, yielding the highest image SNR among the three configurations. These findings support inductive FSS shielding as a compact and practical approach for portable MRI, with improved gradient transparency as an additional benefit.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.