Optimization of a Locally Shielded 16-Channel Transmit Array for Human Brain MRI at 10.5 T via Electromagnetic/Circuit Co-Simulation and Parallel Transmission Strategies
Mengxuan Zheng, Xin Shao, Matt Waks, Mazin M. Mustafa, Kamil Ugurbil, Xiaoping Wu, Xiaotong Zhang, Alireza Sadeghi-Tarakameh
Key points
From the paper's conclusion- In this work, we sought to optimize a locally shielded 16-channel transmit RF array for human neuroimaging at 10.5 T using electromagnetic/circuit co-simulation combined with parallel-transmission strategies.
Abstract
Modern neuroimaging and neuroscience increasingly demand high spatial and temporal resolution, driving the use of ultrahigh-field (7 T and above) MRI for improved signal-to-noise ratio. Realizing this advantage depends critically on advanced radiofrequency (RF) and gradient technologies, particularly multi-channel transmit arrays and high-performance head-gradient systems. In integrated systems, a local RF shield must provide sufficient field confinement while preserving transmit performance and minimizing sensitivity to nearby conductive structures. In this work, we sought to optimize a locally shielded 16-channel transmit RF array for human neuroimaging at 10.5 T using electromagnetic/circuit co-simulation combined with parallel-transmission strategies. Candidate transmit-array and RF-shield configurations were evaluated before and after gradient-shield integration to identify a configuration with suitable RF transmit and shielding performance for the integrated system. The results show that no single standalone metric, including static RF shimming performance, dynamic parallel transmission (pTx) performance, or external-field leakage, reliably predicts robustness after gradient-shield integration. Progressive screening identified the aligned loop-folded dipole (LFD) array combined with a compact cylindrical shield featuring a conical top and three windows (160-mm radius and 70-mm window length) as the final configuration. Without RF-circuit retuning, the selected design maintained external-field leakage below 1% and after gradient shield integration achieved whole-brain excitation coefficient of variation (CoV) of approximately 27% with static RF shimming and 9% with dynamic pTx. Experimental validation of the unshielded LFD array showed good agreement between simulated and measured B1+ efficiency distributions, with a normalized root-mean-square error (NRMSE) of approximately 8%.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Design of an 8‐Channel Transmit 32‐Channel Receive 11.7T Head Coil and Evaluation of SNR GainsMagnetic Resonance in Medicine · 2026
- Detailed Consideration of a Novel Meandered Dipole Array for Magnetic Resonance Imaging of the Head at 3 Tesla with Low Radiofrequency Power DepositionSensors · 2026
- Per‐Channel Complex B1+ Mapping for Parallel Transmit Ultra‐High Field MRINMR in Biomedicine · 2026
- Optimising 7T-fMRI for Imaging Regions of Magnetic Susceptibility.Human brain mapping · 2026
- Design, fabrication and measurement of a force-balanced strong gradient magnetic field assembly for a cryogen-free 7 T / 160 mm MRI superconducting magnet.Magnetic resonance imaging · 2026
- Radio-frequency pulse design in local rotating frame in magnetic resonance imaging.Magnetic resonance imaging · 2026