Journal of the mechanical behavior of biomedical materials

Effects of CBCT acquisition protocols and metal artifact reduction on gray-value measurements of dental restorative materials.

İkbal Esra Pehlivan, Sema Kaya

Department of Restorative Dentistry, Faculty of Dentistry, Van Yuzuncu Yil University, Van, Turkey. Electronic address: ikbalpehlivan@yyu.edu.tr. · Department of Oral and Maxillofacial Radiology, Faculty of Dentistry, Van Yuzuncu Yil University, Van, Turkey. Electronic address: semakaya@yyu.edu.tr.

Key points

From the paper's conclusion
  • The observed ROI gray-value differences were specific to the evaluated phantom setup and acquisition and reconstruction conditions. These measurements do not establish changes in artifact severity or diagnostic accuracy.

Abstract

BACKGROUND: Cone-beam computed tomography (CBCT) imaging is frequently affected by artifacts caused by dental restorative materials, which compromise gray-value accuracy and diagnostic reliability. This study evaluated mean ROI gray values in phantom teeth restored with glass ionomer or composite resin under different CBCT acquisition protocols, field-of-view settings, and metal artifact reduction (MAR) conditions. METHODS: Twenty standardized three-dimensional printed phantom teeth were restored with glass ionomer (n = 10) or composite resin (n = 10). Each specimen was evaluated under three CBCT acquisition protocols (2 mA-0.30 mm, 5 mA-0.30 mm, and 6.3 mA-0.15 mm), two field-of-view settings, and MAR-on and MAR-off reconstruction conditions. A total of 960 region-of-interest (ROI) gray-value measurements were obtained. Data were analyzed using a linear mixed-effects model with a specimen-level random intercept, including MAR × acquisition protocol and MAR × restorative material interactions. Tooth-level paired comparisons were performed with Holm adjustment for multiple testing. RESULTS: MAR-associated gray-value changes differed significantly across acquisition protocols (Wald χ2(2) = 1057.29, p < 0.001), and a significant MAR × restorative material interaction was identified (p < 0.001). In tooth-level paired analyses, significant MAR-associated reductions were observed only under the 6.3 mA-0.15 mm protocol (mean difference: -25.985; 95% CI: -32.188 to -19.781; Holm-adjusted p < 0.001). The reduction was greater for composite resin (-37.314) than for glass ionomer (-14.655). Mandibular field of view was associated with lower gray values (p = 0.014), whereas section orientation and ROI location showed no significant main effects (p > 0.05). CONCLUSIONS: The observed ROI gray-value differences were specific to the evaluated phantom setup and acquisition and reconstruction conditions. These measurements do not establish changes in artifact severity or diagnostic accuracy. CLINICAL SIGNIFICANCE: Protocol- and MAR-associated gray-value changes may have implications for image interpretation, but their diagnostic relevance requires validation in anatomically representative models or clinical studies.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers