Effects of different CBCT acquisition protocols on radiation dose and observer agreement in the assessment of impacted mandibular third molars.
Serhat Efeoglu, Hakan Eren, Rana Nalcaci
Ankara University Faculty of Dentistry and Oral Maxillofacial Radiology, Mevlana Bulvarı, Emniyet Mahallesi, 19/1, Ankara, Yenimahalle, Türkiye. serhat.1996@hotmail.com. · Canakkale Onsekiz Mart University Faculty of Dentistry/ Oral Maxillofacial Radiology, Canakkale, Türkiye.
Key points
From the paper's conclusion- Limited FOV and optimized dose settings reduced radiation exposure while preserving observer agreement, supporting ALADAIP (As Low As Diagnostically Acceptable being Indication-oriented and Patient-specific) principles.
Abstract
This study aimed to evaluate the effects of different CBCT (Cone-beam computed tomography) protocols defined by device type, field of view (FOV), and dose settings on organ-specific radiation doses and inter-/intra-observer agreement in the assessment of impacted mandibular third molars. A total of 168 CBCT scans (NewTom 7G and NewTom Go) were retrospectively analyzed. Each device included six protocols (different FOVs and dose levels: low, regular, best). Organ and effective doses were calculated using the Monte Carlo PCXMC 2.0 Rotation software. Three observers independently assessed five anatomical parameters related to surgical risk. Inter- and intra-observer agreement was analyzed using Fleiss' and Cohen's kappa. Effective doses ranged from 8.65 to 376.96 µSv, a more than 40-fold difference between the lowest- and highest-dose protocols. NewTom Go protocols with limited FOV and low/regular doses yielded the lowest effective doses without a corresponding loss of observer agreement. Observer agreement was substantial to almost perfect for most parameters (kappa: 0.541-0.915), although agreement for periodontal ligament visibility and root resorption was partly driven by shared "not visible"/"uncertain" ratings rather than confident structure identification. As no clinical or surgical reference standard was used, these findings reflect observer agreement and diagnostic interpretability rather than confirmed diagnostic accuracy. Limited FOV and optimized dose settings reduced radiation exposure while preserving observer agreement, supporting ALADAIP (As Low As Diagnostically Acceptable being Indication-oriented and Patient-specific) principles.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Comparing reproducibility among observers for low-, standard-, and high-dose protocols of CBCT assessing anatomical landmarks and pathology related to third molars: An ex vivo study.Oral surgery, oral medicine, oral pathology and oral radiology · 2026
- Within-Device Exposure-Setting and Voxel-Size Changes in CBCT Alveolar Bone Measurement for Implant Planning: A Systematic Review.Tomography (Ann Arbor, Mich.) · 2026
- Diagnostic accuracy of three CBCT protocols and Metal Artifact Reduction tool for detecting the second canal of mandibular incisors adjacent to a zirconia implant.Dento maxillo facial radiology · 2026
- Selective Cone-Beam CT Referral for Mandibular Third Molars Based on Panoramic Risk Signs: A Retrospective Cross-Sectional Diagnostic StudyDentistry Journal · 2026
- Evaluating cone-beam computed tomography vs. intraoral radiography for the detection of endodontal disease in dogs: diagnostic performance and inter-rater agreement.Frontiers in veterinary science · 2026
- Diagnostic Reference Levels for Dental Cone Beam Computed Tomography: A Critical Appraisal of Global Progress and Remaining Gaps.Cureus · 2026