Comparison of Cone-Beam CT, Photon-Counting Detector CT and Micro-CT for Imaging Bone Graft Biomaterials and Three-Dimensional Printed Calcium Phosphate Scaffolds: An Ex Vivo Study.
Julien Guerrero, Paula Korn, Adib Al-Haj Husain, Victor Mergen, Tristan T Demmert, Hatem Alkadhi, Ana Pérez Domínguez, Richard F Richter,
Center for Dental Medicine, Oral Biotechnology & Bioengineering, University of Zurich, Plattenstrasse 11, 8032 Zurich, Switzerland. · Department of Oral and Maxillofacial Surgery, Faculty of Medicine Carl Gustav Carus, University Hospital Carl Gustav Carus, Technische Universität Dresden, 1067 Dresden, Germany.
Key points
From the paper's conclusion- Conclusions: In this exploratory ex vivo study, the evaluated PCD-CT protocols provided favorable qualitative visualization of several bone graft materials and three-dimensional printed calcium phosphate scaffolds compared with the evaluated CBCT protocol, particularly in the presence of titanium cages.
Abstract
Background/Objectives: Cone-beam computed tomography (CBCT) is widely used in dentomaxillofacial imaging for evaluating bone grafts and implant-related procedures; however, image quality is often compromised by metal-induced artefacts. This study compared CBCT, photon-counting detector CT (PCD-CT), and micro-computed tomography (µCT) for imaging clinically relevant bone graft materials and three-dimensional printed calcium phosphate scaffolds, focusing on biomaterial visualisation and titanium artefact reduction. Methods: Four bovine mandibles were prepared with standardized cylindrical defects containing eight bone substitute materials and 3D-printed calcium phosphate scaffolds, with or without surrounding titanium cages. Specimens were imaged using CBCT and PCD-CT with low-dose and maximum-dose ultra-high-resolution acquisition protocols, and µCT. Two readers independently assessed image interpretability, biomaterial visualisation, and metal artefact severity using a five-point scoring system. Inter-reader agreement was evaluated using weighted Cohen's kappa. Results: µCT provided the highest spatial resolution and served as the ex vivo reference standard. Under the acquisition and reconstruction protocols evaluated, PCD-CT provided favorable visualization of scaffold architecture and reduced the visual impact of titanium-induced artefacts compared with CBCT. High-energy virtual monoenergetic reconstructions (≥130 keV) further reduced beam-hardening and streak artefacts, although the optimal reconstruction energy may depend on the biomaterial and diagnostic task. Without titanium cages, PCD-CT and µCT achieved higher image interpretability scores than CBCT. With titanium cages, CBCT frequently became non-diagnostic, whereas PCD-CT maintained good visualisation of most biomaterials. Inter-reader agreement ranged from substantial to almost perfect. Conclusions: In this exploratory ex vivo study, the evaluated PCD-CT protocols provided favorable qualitative visualization of several bone graft materials and three-dimensional printed calcium phosphate scaffolds compared with the evaluated CBCT protocol, particularly in the presence of titanium cages. However, the protocols were not dose-matched, and differences in acquisition and reconstruction parameters prevent attributing the observed findings exclusively to detector technology. The study does not establish PCD-CT superiority at equivalent radiation doses or demonstrate clinical performance in longitudinal bone-regeneration monitoring. Further dose-matched studies incorporating objective image-quality measurements and clinical validation are required.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Diagnostic performance of clinical photon-counting CT in the assessment of peri-implantitis - an ex vivo diagnostic accuracy study.International journal of implant dentistry · 2026
- Photon-Counting CT in Oral and Maxillofacial Imaging: Clinical Applications, Dose Efficiency, and Future Perspectives.Diagnostics (Basel, Switzerland) · 2026
- Postoperative assessment of fracture reduction and osteosynthesis materials using photon-counting detector CT in maxillofacial trauma – a pilot studyOral and Maxillofacial Surgery · 2025
- Photon-Counting Detector CT for Imaging Around Orthopedic and Spinal Metallic Implants: A Systematic Review.Cureus · 2026
- Artifacts of orthodontic materials on cone beam computed tomography images.BMC oral health · 2026
- Photon-Counting CT in Musculoskeletal Imaging: Emerging Applications and Clinical Implications.Canadian Association of Radiologists journal = Journal l'Association canadienne des radiologistes · 2026