Influence of Clinical Crown Length and Alignment Method on Accuracy and Buccal Occlusal Scanning Time in Maxillo-Mandibular Registration Using an Intraoral Scanner.
Chunxiao Jin, Yaping Guo, Liya Ma, Xinjie Cai, Cui Huang
State Key Laboratory of Oral & Maxillofacial Reconstruction and Regeneration, Key Laboratory of Oral Biomedicine Ministry of Education, Hubei Key Laboratory of Stomatology, School & Hospital of Stomatology, Wuhan University; Department of Stomatology, Peking Union Medical College Hospital, Chinese Academy of Medical Sciences and Peking Union Medical College, Beijing, 100730, China. · State Key Laboratory of Oral & Maxillofacial Reconstruction and Regeneration, Key Laboratory of Oral Biomedicine Ministry of Education, Hubei Key Laboratory of Stomatology, School & Hospital of Stomatology, Wuhan University.
Key points
From the paper's conclusion- Reduced clinical crown length may compromise the overall trueness and efficiency of digital maxillomandibular registration, although regional deviations do not necessarily increase uniformly across the dental arch. Manual alignment achieves comparable accuracy with shorter scanning time, particularly in dentitions with markedly reduced clinical crown exposure.
Abstract
OBJECTIVES: To evaluate how clinical crown length and alignment method affect the accuracy and buccal occlusal scan time required for digital maxillomandibular relationship registration. METHODS: Four maxillary-mandibular cast pairs with clinical crown lengths of 2, 4, 6, or 8 mm (G2-G8) were fabricated. Each pair was mounted in maximum intercuspation and scanned with a desktop scanner to generate reference models. Corresponding scans were obtained with an intraoral scanner (TRIOS 5). Ten occlusal records per group were registered using automatic and manual alignment. For G2, buccal occlusal scans of shorter durations that did not trigger automatic registration were additionally performed and manually aligned. Trueness was assessed by measuring positional deviations of reference cubes between test and reference datasets. RESULTS: Crown length significantly affected registration trueness, which decreased as clinical crown length decreased (p < 0.05), whereas automatic and manual alignment showed comparable accuracy within all groups. Shorter crowns (G2 and G4) produced greater dentition distance deviations than longer crowns (G6 and G8) (p < 0.05). The scan time required to trigger automatic registration increased as crown length decreased, reaching up to five times longer in G2 than G8 (p < 0.05). In G2, shortening the buccal occlusal scan duration and subsequently performing manual alignment did not significantly affect registration accuracy (p > 0.05). CONCLUSION: Reduced clinical crown length may compromise the overall trueness and efficiency of digital maxillomandibular registration, although regional deviations do not necessarily increase uniformly across the dental arch. Manual alignment achieves comparable accuracy with shorter scanning time, particularly in dentitions with markedly reduced clinical crown exposure. CLINICAL SIGNIFICANCE: For patients with short clinical crowns, manual alignment may shorten buccal occlusal scanning time without compromising registration accuracy, potentially reducing the risk of artifacts related to patient movement.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Comparative Evaluation of Tooth Type and Pulp Chamber Depth on the Accuracy of Endocrown Scans Acquired Using Various Intraoral Scanners.Journal of esthetic and restorative dentistry : official publication of the American Academy of Esthetic Dentistry ... [et al.] · 2026
- The impact of different cavity design parameters on intraoral scanner accuracy in endocrown restorations.American journal of dentistry · 2026
- Scanner-Dependent Influence of Palatal Surface Acquisition on Trueness, Precision and Transverse Arch Reproduction During Maxillary Complete-Arch Intraoral Scanning.Journal of dentistry · 2026
- Trueness of sequential and simultaneous capture strategies for adjacent implant scan bodies.Journal of dentistry · 2026
- Accuracy of using novel bite registration caps to record the intercuspal position by intraoral scanning for fixed implant prostheses with different edentulous span lengths.Journal of dentistry · 2026
- Accuracy of complete arch implant intraoral scans: A systematic review of different technique categories compared with the conventional splinted open tray method.The Journal of prosthetic dentistry · 2026