Biomechanical and fatigue analysis of a porous titanium total temporomandibular joint prosthesis with different mandibular component lengths and designs.

Aydın Onur Gerçek, Salih Eren Meral, Hakan Hıfzı Tüz

Department of Oral and Maxillofacial Surgery, Faculty of Dentistry, Hacettepe University, Ankara, Turkey. Electronic address: aogercek@gmail.com. · Department of Oral and Maxillofacial Surgery, Faculty of Dentistry, Hacettepe University, Ankara, Turkey. Electronic address: erenmeral@gmail.com.

Key points

From the paper's conclusion
  • Ramus stress decreased with increasing filling ratio, ranging from 363.5 MPa (TMJR 25% Voronoi) to 71.3 MPa (eTMJR solid); all designs showed infinite fatigue life, with safety factors of 1.5-7.8.

Abstract

This study evaluated the biomechanical behavior and fatigue lifetime of Voronoi lattice-structured ramus components for total temporomandibular joint replacement (TMJR) in mini (mini-TMJR), standard (TMJR), and extended (eTMJR) configurations based on the Walter Lorenz (Biomet) system, using finite element analysis. Three-dimensional models were derived from Visible Human Project computed tomography data. For each ramus length, four designs were analyzed: Voronoi architectures with 25%, 50%, and 75% filling ratios, and a fully solid design, under masticatory forces simulating unilateral left molar clenching. Von Mises stress, principal bone stress, and fatigue safety factors were evaluated. Ramus stress decreased with increasing filling ratio, ranging from 363.5 MPa (TMJR 25% Voronoi) to 71.3 MPa (eTMJR solid); all designs showed infinite fatigue life, with safety factors of 1.5-7.8. Fossa and ultra-high-molecular-weight polyethylene (UHMWPE) stresses increased with filling ratio, with UHMWPE stress exceeding the polyethylene yield strength in all models, while bone stress remained within physiological limits. Low filling ratio designs transferred more load to bone, which may reduce stress shielding, whereas the eTMJR solid design showed the lowest ramus stress and highest safety factor. Voronoi-latticed ramus components showed acceptable biomechanical performance under the simulated conditions, though long-term biological and clinical validation is required.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers

Biomechanical and fatigue analysis of a porous titanium total temporomandibular joint prosthesis with different mandibular component lengths and designs — Science Skim