Journal of chromatography. B, Analytical technologies in the biomedical and life sciences

An automated column-switching LC-MS/MS method using immobilized trypsin for the rapid quantification of pembrolizumab in human serum.

Kaito Shibata, Honoka Shimizu, Kazuki Abe, Junichi Kawakami, Takafumi Naito

Department of Pharmacy, Shinshu University Hospital, Matsumoto, Nagano, Japan; Department of Pharmacy, Central Japan International Medical Center, Minokamo, Gifu, Japan. · Department of Hospital Pharmacy, Hamamatsu University School of Medicine, Hamamatsu, Japan.

Key points

From the paper's conclusion
  • In conclusion, this automated column-switching LC-MS/MS method shows acceptable analytical performance and may support therapeutic drug monitoring of serum pembrolizumab.

Abstract

An automated column-switching liquid chromatography coupled to tandem mass spectrometry (LC-MS/MS) method was developed for the rapid quantification of pembrolizumab in human serum. The workflow integrates rapid digestion using an immobilized trypsin with automated online desalting. The column-switching system with an online solid-phase extraction column concentrated the signature peptide and removed matrix components prior to MS detection. Chromatographic analysis with a stepwise gradient was completed in 8 min, reducing the per-sample workflow to approximately 45 min. No chromatographic interference was observed at the retention time of the signature peptide or the stable isotope-labeled internal standard in blank serum digests. The dynamic range of the calibration curve for serum pembrolizumab was 1-140 μg/mL. The validation demonstrated negligible matrix effects and minimal carryover. The intra- and inter-day accuracies and imprecisions were 93.2%-99.9% and less than 7.6%, respectively. Pre-dose serum levels of pembrolizumab ranged from 6.8 to 69.0 μg/mL in 25 patients with urothelial carcinoma. Comparison of the LC-MS/MS method with the ELISA method showed no clear systematic bias, although substantial variability was observed in individual differences between the two methods. In conclusion, this automated column-switching LC-MS/MS method shows acceptable analytical performance and may support therapeutic drug monitoring of serum pembrolizumab.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers