Journal of pharmaceutical and biomedical analysis

LC-MS/MS method for quantification of dupilumab in human serum using Protein G enrichment.

Kun Yang, Jianhua Wu, Chengming Yang, Liang Liu

Department of Pharmacy, Zhongnan Hospital of Wuhan University, Wuhan 430071, China. · Department of Pharmacy, The First Affiliated Hospital of Yangtze University, Jingzhou, Hubei 434000, China.

Key points

From the paper's conclusion
  • Specifically, dupilumab was enriched from 50 µL serum using Protein G magnetic beads, followed by denaturation, reduction, alkylation, and tryptic digestion.

Abstract

Dupilumab, a monoclonal antibody targeting IL-4Rα, is used for the treatment of type 2 inflammatory diseases. Recent studies have revealed a nonlinear relationship between its serum concentration and clinical response, particularly a "biological tipping point" during dose interval prolongation, highlighting the need for individualized therapeutic drug monitoring (TDM). In this study, a robust liquid chromatography-tandem mass spectrometry (LC-MS/MS) method for dupilumab quantification in human serum was developed and validated. Specifically, dupilumab was enriched from 50 µL serum using Protein G magnetic beads, followed by denaturation, reduction, alkylation, and tryptic digestion. A surrogate peptide SGQSPQLLIYLGSNR was selected for quantification, and its stable isotope-labeled (SIL) analog served as an internal standard. Separation was performed on an Acquity UPLC BEH C8 column (50 mm × 2.1 mm, 1.7 µm) with detection using positive electrospray ionization in multiple reaction monitoring (MRM) mode. Excellent linearity was achieved from 1.00 to 400 µg/mL (r² > 0.99), with a lower limit of quantification of 1.00 µg/mL. Intra- and inter-run precision (%CV) and accuracy (%Bias) were within 9.27%. The validated method was successfully applied to analyze clinical serum samples. The established LC-MS/MS method provides a reliable alternative to immunoassays, with a wide dynamic range suitable for TDM and personalized dose interval adjustments in patients receiving dupilumab.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers