Off-Pump Retrieval of Patent Ductus Arteriosus Closure Device Causing Partial Aortic Obstruction.
Paighton C Miller, Jacob R Miller, Pirooz Eghtesady
Section of Pediatric Cardiothoracic Surgery, Department of Surgery, Washington University School of Medicine, St. Louis Children's Hospital, St. Louis, Missouri, USA. · Section of Pediatric Cardiothoracic Surgery, Department of Surgery, Washington University School of Medicine, St. Louis Children's Hospital, St. Louis, Missouri, USA. Electronic address: eghtesady670@wustl.edu.
Key points
From the paper's conclusion- Off-pump device retrieval for partial aortic obstruction, while addressing the ductus arteriosus, can be accomplished safely via median sternotomy.
Abstract
BACKGROUND: Transcatheter devices for closure of patent ductus arteriosus (PDA) in preterm and low-birth-weight infants have become increasingly utilized in practice. Device migration is a rare but known risk. CASE SUMMARY: An 8-week-old preterm male with moderate PDA underwent attempted catheter-based closure complicated by partial migration and aortic obstruction. Off-pump surgical retrieval was successfully performed with minimal duration of distal flow occlusion. DISCUSSION: Surgery may be required for transcatheter device migration when transcatheter interventions for retrieval fail, in which case, patient age and size contribute greatly to complexity and risk of cardiopulmonary bypass. Further, techniques regarding device retrieval vary greatly based on location of migration and site of obstruction. Off-pump intervention, as described here, is a safe, feasible, and perhaps optimal approach for these premature babies with multiple comorbidities in select circumstances. TAKE-HOME MESSAGES: Device complications of transcatheter PDA closure require thoughtful intervention. Off-pump device retrieval for partial aortic obstruction, while addressing the ductus arteriosus, can be accomplished safely via median sternotomy.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Transcatheter Closure of Residual Shunt Using the Coil-in-Plug Method After Amplatzer Vascular Plug Implantation for Tubular Patent Ductus Arteriosus in Infants: Two Case Reports.Pediatric cardiology · 2026
- Balloon-assisted TorqVue TM delivery catheter advancement for residual patent ductus arteriosus closure in a postsurgical infant with Edwards syndrome.Cardiology in the young · 2026
- Pulmonary Artery Dissection in Unrepaired Patent Ductus Arteriosus: Hemodynamic-Guided Ductal Closure.JACC. Case reports · 2026
- Late Percutaneous Retrieval of a Septal Occluder to Facilitate Left Atrial Access.JACC. Case reports · 2026
- Novel Transcatheter Device Use in Patent Ductus Arteriosus Closure in an 800-g Infant.JACC. Case reports · 2026
- Suspected Infective Endocarditis in a Premature Neonate After Patent Ductus Arteriosus Occluder Placement: A Case Report.Case reports in pediatrics · 2026