Transcatheter Closure of Residual Shunt Using the Coil-in-Plug Method After Amplatzer Vascular Plug Implantation for Tubular Patent Ductus Arteriosus in Infants: Two Case Reports.
Mitsuhiro Fujino, Kae Nakamura, Takeshi Sasaki, Yuki Kawasaki, Hisashi Sugiyama
Department of Pediatric Cardiology and Electrophysiology, Osaka City General Hospital, 2-13-22 Miyakojima-hondori, Miyakojima-ku, Osaka, 534-0021, Japan. m-fujino@med.osakacity-hp.or.jp. · Department of Pediatric Cardiology and Electrophysiology, Osaka City General Hospital, 2-13-22 Miyakojima-hondori, Miyakojima-ku, Osaka, 534-0021, Japan.
Key points
From the paper's conclusion- We report two infants with a large PDA and slight aortic isthmus narrowing who underwent implantation of an 8-mm AVP I.
Abstract
Transcatheter closure of a wide, short tubular patent ductus arteriosus (PDA) in infants is challenging because devices may protrude into adjacent vessels. The Amplatzer Vascular Plug I (AVP I) has no retention disk and suits tubular morphology, but its lack of internal fabric increases the risk of residual shunting. The coil-in-plug (CIP) method, in which microcoils are deployed within a previously implanted AVP I, has been proposed to overcome this limitation. We report two infants with a large PDA and slight aortic isthmus narrowing who underwent implantation of an 8-mm AVP I. Persistent residual shunting prompted staged CIP procedures two months later, allowing time for endothelialization. Detachable microcoils were delivered through a microcatheter, achieving complete occlusion without complications. CIP may be a rescue option for residual shunting after device placement in infants with challenging tubular PDAs, provided sufficient time has elapsed for device fixation.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Balloon-assisted TorqVue TM delivery catheter advancement for residual patent ductus arteriosus closure in a postsurgical infant with Edwards syndrome.Cardiology in the young · 2026
- Novel Transcatheter Device Use in Patent Ductus Arteriosus Closure in an 800-g Infant.JACC. Case reports · 2026
- Off-Pump Retrieval of Patent Ductus Arteriosus Closure Device Causing Partial Aortic Obstruction.JACC. Case reports · 2026
- Clinical outcomes and midterm durability of the preloading coil-in-plug technique in various arterial territories.CVIR endovascular · 2026
- Plug-Anchored Coil Embolization of a Paravalvular Aortic Pseudoaneurysm After TAVR With BASILICA.JACC. Case reports · 2026
- Making sense of the evidence: designing randomised controlled trials for preterm infants with high-shunt volume patent ductus arteriosusArchives of Disease in Childhood Fetal & Neonatal · 2026