JACC. Case reports

Single-Access Retrieval of an Entrapped Rotational Atherectomy Burr Using Parallel-Wire STAR Technique.

Anwar Hussain, Ahmed Elsherif, Sophia Khattak, Farhan Shahid, Sohail Q Khan

Department of Interventional Cardiology, Queen Elizabeth Hospital, University Hospitals Birmingham NHS Foundation Trust, Birmingham, United Kingdom. · Department of Interventional Cardiology, Queen Elizabeth Hospital, University Hospitals Birmingham NHS Foundation Trust, Birmingham, United Kingdom; Institute of Cardiovascular Sciences, University of Birmingham, Birmingham, United Kingdom. Electronic address: s.khattak@bham.ac.uk.

Key points

From the paper's conclusion
  • Successful retrieval of the burr was achieved via a single 7-F access by cutting the burr and extracting the drive shaft sheath, followed by subintimal tracking and re-entry (STAR) with a parallel wire, then external crushing to modify the calcified plaque.

Abstract

BACKGROUND: Rotational atherectomy facilitates percutaneous coronary intervention in heavily calcified coronary lesions. Burr entrapment occurs in <1% of procedures, but it may cause vessel occlusion and myocardial ischemia. CASE SUMMARY: An 81-year-old woman with recurrent angina underwent rotational atherectomy-assisted percutaneous coronary intervention to a heavily calcified right coronary artery. The 1.5-mm burr became entrapped, resulting in inferior ST-segment elevation myocardial infarction. Successful retrieval of the burr was achieved via a single 7-F access by cutting the burr and extracting the drive shaft sheath, followed by subintimal tracking and re-entry (STAR) with a parallel wire, then external crushing to modify the calcified plaque. DISCUSSION: This case demonstrates parallel-wire balloon-assisted plaque modification as an effective percutaneous bailout strategy for an entrapped rotablation burr. TAKE-HOME MESSAGE: Subintimal tracking and re-entry (STAR) with a parallel wire and external plaque crushing by a balloon is feasible from a single 7-F sheath by cutting the burr and extracting the drive-shaft sheath.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers

Single-Access Retrieval of an Entrapped Rotational Atherectomy Burr Using Parallel-Wire STAR Technique — Science Skim