Rotablator Burr Entrapment: A Systematic Review of 215 Published Cases, Mechanisms, and Management.
Evangelos Kolotouros, Vasileios Lerakis, Marios Tsourous, Theodoros Vichos, Pedro E P Carvalho, Ahmet Akyel, Parth Desai, Michaella Alexandrou,
Minneapolis Heart Institute and Minneapolis Heart Institute Foundation, Abbott Northwestern Hospital, Minneapolis, Minnesota, USA. · Athens Naval Hospital, Greece.
Key points
From the paper's conclusion- Burr entrapment was reported in approximately 0.6% of rotational atherectomy procedures. Our findings support a mechanism-based, stepwise approach to burr retrieval. Percutaneous treatment is successful in most contemporary cases.
Abstract
BACKGROUND: Burr entrapment is a rare but potentially life-threatening complication of rotational atherectomy. We reviewed all related publications, classified the mechanisms of entrapment, and summarized management strategies. METHODS: PubMed, Embase, and Google Scholar were searched for English-language publications reporting Rotablator burr entrapment or related retained-burr or device-loss events from inception through August 31, 2026. RESULTS: One hundred fourteen publications were included, representing 215 unique eligible cases after exclusion of one overlapping case. The mechanism was reported in 117 cases, management in 162, and final outcome in 160. Among 34 studies reporting the total number of rotational atherectomy procedures, burr entrapment occurred in 118 of 19,672 procedures (0.6%). The most common classified mechanism was intralesional embedding or wedging (32.5%), followed by the Kokeshi phenomenon (30.8%), stent-related entrapment (26.5%), and primary driveshaft fracture or burr disconnection (10.3%). Among the 162 cases with available management data, 86.4% were managed percutaneously, including 3.7% with intentional burr retention or exclusion (crushing or jailing), and 13.6% required surgical bailout. Management strategies shifted over time, with surgical bailout declining from 66.7% before 2010 to 22.7% in 2010-2019 and 5.5% in 2020-2026. Procedural outcomes differed according to mechanism. Successful burr resolution with completed or otherwise acceptable PCI was most frequent in Kokeshi-type entrapment (80.6%), whereas stent-related entrapment had the highest surgical bailout rate (29.0%). CONCLUSIONS: Burr entrapment was reported in approximately 0.6% of rotational atherectomy procedures. Our findings support a mechanism-based, stepwise approach to burr retrieval. Percutaneous treatment is successful in most contemporary cases.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Single-Access Retrieval of an Entrapped Rotational Atherectomy Burr Using Parallel-Wire STAR Technique.JACC. Case reports · 2026
- Wire-entanglement retrieval of a detached rotational atherectomy burr after complete wire disengagement: a case reportEuropean Heart Journal - Case Reports · 2026
- Successful Bailout of Guidewire Entrapment Within a Coronary Stent Using a Tornus Penetration Catheter: A Case Report.Cureus · 2026
- Guide Extension and Balloon-Assisted Retrieval of an Entrapped Embolic Protection Device Within a Coronary Stent: A Case Report and In Vitro Analysis.Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions · 2026
- Structural degradation of a rotational atherectomy burr following treatment of a massively calcified coronary lesion: Insights from scanning electron microscopy.Annales de cardiologie et d'angeiologie · 2026
- Gutter Rotation Technique for Percutaneous Ipsilateral Retrieval of a Ruptured SAPIEN-3 Balloon.Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions · 2026