JACC. Case reports

Left Atrial Appendage Occlusion and Percutaneous Coumadin Ridge Mass Exclusion: Two Targets, One Device.

Alexander Ottersbach, Adam Bisharat, Dimitri Sherev, Nassir A Azimi

Azimi Cardiovascular Institute, La Mesa, California, USA. · Sharp Grossmont Hospital, La Mesa, California, USA; Sherev Heart and Vascular, El Cajon, California, USA.

Case report

Key points

From the paper's conclusion
  • This is to our knowledge the first reported case of intentional simultaneous LAA occlusion and percutaneous mass exclusion, demonstrating a novel use of device geometry to manage dual embolic risk.

Abstract

BACKGROUND: Mobile cardiac masses on the coumadin ridge are rare and carry significant embolic risk, compounded by concurrent atrial fibrillation. Surgical excision may be prohibitive in patients with high bleeding risk. CASE SUMMARY: A 70-year-old man with paroxysmal atrial fibrillation, a HAS-BLED score of 4, and aversion to anticoagulation was referred for left atrial appendage (LAA) closure. Imaging revealed an incidental 0.56-cm mobile coumadin ridge mass consistent with papillary fibroelastoma. The Amplatzer Amulet was selected for its larger disc geometry to simultaneously occlude the LAA and exclude the mass, with follow-up confirming complete mass exclusion and stability at over 2 years. DISCUSSION: This is to our knowledge the first reported case of intentional simultaneous LAA occlusion and percutaneous mass exclusion, demonstrating a novel use of device geometry to manage dual embolic risk. TAKE-HOME MESSAGE: Concurrent LAA and coumadin ridge mass risk can be addressed in a single procedure through deliberate device selection using the Amulet's larger disc.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers

Left Atrial Appendage Occlusion and Percutaneous Coumadin Ridge Mass Exclusion: Two Targets, One Device — Science Skim