European heart journal. Case reports

Persistent peri-device leak following left atrial appendage occlusion treated with hybrid coil- and plug-based closure: a case report.

Ardashir Ibrahim Sulaiman, Jens Trøan, Martin Kirk Christensen, Lisette Okkels Jensen, Mikkel Hougaard

Department of Cardiology, Odense University Hospital, J.B. Winsløwsvej 4, Odense C 5000, Denmark.

Case report

Key points

From the paper's conclusion
  • This case provides insight into the potential role of hybrid coil- and plug-based strategies in enhancing occlusion but also highlights that anticoagulation remains necessary when full occlusion is not obtained.

Abstract

BACKGROUND: Left atrial appendage occlusion (LAAO) is an emerging alternative to long-term anticoagulation in patients with atrial fibrillation, but peri-device leaks remain a clinical challenge. CASE SUMMARY: A 63-year-old man with permanent atrial fibrillation, heart failure, a prior ischaemic stroke history was referred for LAAO due to recurrent anticoagulation-related bleeding. Pre-procedural computer tomography (CT) showed a markedly enlarged appendage suitable for a 35-mm WATCHMAN FLX, which was implanted under intracardiac echocardiography guidance, with subsequent transition to aspirin. Three months later, CT revealed a tilted device with a peri-skirt leak; an ischaemic stroke had occurred in the interim, and anticoagulation was resumed. An Amplatzer Vascular Plug II was implanted but follow-up CT demonstrated persistent contrast passage and more but incomplete occlusion. Three months later, the patient experienced a second stroke, and CT confirmed ongoing peri-device flow. A hybrid leak-closure procedure was performed with coils in the appendage and within the device, followed by a 12-mm plug at the leak channel, achieving near-complete occlusion. At 7-month follow-up, the patient remained on rivaroxaban therapy, briefly interrupted due to haematuria associated with nephrolithiasis. DISCUSSION: Most cases of peri-device leak after LAAO are related to suboptimal device alignment, incomplete endothelialization, or persistent peri-skirt channels. In our case, device tilt with a peri-skirt tract persisted despite initial plug-based closure. This case provides insight into the potential role of hybrid coil- and plug-based strategies in enhancing occlusion but also highlights that anticoagulation remains necessary when full occlusion is not obtained.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers