Persistent peri-device leak following left atrial appendage occlusion treated with hybrid coil- and plug-based closure: a case report.
Ardashir Ibrahim Sulaiman, Jens Trøan, Martin Kirk Christensen, Lisette Okkels Jensen, Mikkel Hougaard
Department of Cardiology, Odense University Hospital, J.B. Winsløwsvej 4, Odense C 5000, Denmark.
Key points
From the paper's conclusion- This case provides insight into the potential role of hybrid coil- and plug-based strategies in enhancing occlusion but also highlights that anticoagulation remains necessary when full occlusion is not obtained.
Abstract
BACKGROUND: Left atrial appendage occlusion (LAAO) is an emerging alternative to long-term anticoagulation in patients with atrial fibrillation, but peri-device leaks remain a clinical challenge. CASE SUMMARY: A 63-year-old man with permanent atrial fibrillation, heart failure, a prior ischaemic stroke history was referred for LAAO due to recurrent anticoagulation-related bleeding. Pre-procedural computer tomography (CT) showed a markedly enlarged appendage suitable for a 35-mm WATCHMAN FLX, which was implanted under intracardiac echocardiography guidance, with subsequent transition to aspirin. Three months later, CT revealed a tilted device with a peri-skirt leak; an ischaemic stroke had occurred in the interim, and anticoagulation was resumed. An Amplatzer Vascular Plug II was implanted but follow-up CT demonstrated persistent contrast passage and more but incomplete occlusion. Three months later, the patient experienced a second stroke, and CT confirmed ongoing peri-device flow. A hybrid leak-closure procedure was performed with coils in the appendage and within the device, followed by a 12-mm plug at the leak channel, achieving near-complete occlusion. At 7-month follow-up, the patient remained on rivaroxaban therapy, briefly interrupted due to haematuria associated with nephrolithiasis. DISCUSSION: Most cases of peri-device leak after LAAO are related to suboptimal device alignment, incomplete endothelialization, or persistent peri-skirt channels. In our case, device tilt with a peri-skirt tract persisted despite initial plug-based closure. This case provides insight into the potential role of hybrid coil- and plug-based strategies in enhancing occlusion but also highlights that anticoagulation remains necessary when full occlusion is not obtained.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Left Atrial Appendage Occlusion with the Watchman FLX Device: A Single-Centre Retrospective Analysis of Procedural and One-Year Clinical Outcomes.Medical sciences (Basel, Switzerland) · 2026
- Left atrial appendage closure: Current status, unresolved issues, and future perspectivesCardiovascular Intervention and Therapeutics · 2025
- Endocardial Left Atrial Appendage Occlusion in Atrial Fibrillation: Device Design, Clinical Evidence, and Postprocedural Antithrombotic Management.Medicina (Kaunas, Lithuania) · 2026
- Left Atrial Appendage Closure Complicated by Spinal Epidural Hematoma and Device‐Related Thrombosis: A Case ReportPacing and Clinical Electrophysiology · 2026
- Left Atrial Appendage Electrical Isolation Reduces Residual Patency After Device Closure in Patients With Persistent Atrial FibrillationJournal of the American Heart Association · 2026
- Epicardial Closure for Peridevice Leaks from Endocardial Left Atrial Appendage Closure - Early LARIAT-4-PDL Registry Results.Heart rhythm · 2026