Device embolization after percutaneous left atrial appendage occlusion: a systematic review and meta-analysis.
Mattia Basile, Miguel A Martín-Arena, Jon Zubiaur-Zamacola, Ignacio Cruz-Gonzalez, Borja Rivero-Santana, Santiago Jiménez-Valero, Daniel Tébar-Márquez, Alfonso Jurado-Román,
Department of Cardiology, La Paz University Hospital, Madrid, Spain. · Centro de Investigación Biomédica en Red de Enfermedades Cardiovasculares (CIBER-CV), Instituto de Salud Carlos III, Madrid, Spain.
Key points
From the paper's conclusion- Device embolization after LAAO is infrequent and has declined substantially over time, likely reflecting technological advances and procedural experience. However, when embolization occurs, it remains associated with high mortality, underscoring the importance of careful device selection and optimized implantation strategies.
Abstract
BACKGROUND AND AIMS: Left atrial appendage occlusion (LAAO) is an established alternative for stroke prevention in patients with atrial fibrillation who are unsuitable for long-term anticoagulation. Although device embolization is rare, it represents a potentially catastrophic complication. Robust contemporary data on its incidence, timing, determinants, and clinical consequences are limited. To provide contemporary, comprehensive estimates of device embolization after percutaneous LAAO and to identify factors associated with embolization risk and related mortality. METHODS: We conducted a systematic review and meta-analysis of studies reporting device embolization after LAAO, as well as embolization-related mortality. Random-effects models were applied to estimate pooled embolization incidence and case-fatality rates. Subgroup analyses were performed by device type and device family. Exploratory meta-regression assessed sources of heterogeneity, including publication year and study sample size. RESULTS: Seventy-three studies, including seventy observational studies and three randomized controlled trials, involving 97 316 patients, were analysed. The pooled embolization incidence was .31% [95% confidence interval (CI) .19%-.50%]. Among studies reporting embolization timing, 80% of events occurred within 24 h of implantation. Larger sample size and more recent publication year were independently associated with lower embolization risk. Device-specific incidence was lowest for Watchman FLX (.02%) and Watchman 2.5 (.12%), and higher for Amplatzer Cardiac Plug (1.02%) and Amulet (.43%). Embolization-related mortality was 10.61% (95% CI 4.98%-21.16%), with no evidence of temporal improvement. CONCLUSIONS: Device embolization after LAAO is infrequent and has declined substantially over time, likely reflecting technological advances and procedural experience. However, when embolization occurs, it remains associated with high mortality, underscoring the importance of careful device selection and optimized implantation strategies.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Delayed Aortic Embolization of a Left Atrial Appendage Occlusion Device: Implications for Risk-Stratified Postprocedural Management.JACC. Case reports · 2026
- Incidence, Predictors, Management, and Clinical Impact of Major Bleeding Following Left Atrial Appendage OcclusionJAMA Cardiology · 2026
- Embolic events after left atrial appendage closure for recurrent embolic events despite anticoagulationEuropean Heart Journal · 2026
- Left Atrial Appendage Occlusion Versus Oral Anticoagulation in Atrial Fibrillation: A Comprehensive Evidence Synthesis.Pacing and clinical electrophysiology : PACE · 2026
- Updated Meta-Analysis of Left Atrial Appendage Closure Versus Oral Anticoagulation to Prevent Stroke in Atrial Fibrillation.Stroke · 2026
- Left Atrial Appendage Occlusion vs Oral Anticoagulation for Stroke Prevention in Atrial Fibrillation: A Meta-Analysis.Pacing and clinical electrophysiology : PACE · 2026