A Perforation Mimic After Temporary Pacing Lead Migration: Surgical Confirmation of Marked Apical Wall Thinning.
Yudai Tanaka, Riku Arai, Kohei Agune, Hikaru Masuda, Koichiro Hori, Yasunari Ebuchi, Atsushi Harada, Keisuke Kojima,
Division of Cardiology, Department of Medicine, Nihon University School of Medicine, Itabashi-ku, Tokyo, Japan. · Division of Cardiology, Department of Medicine, Nihon University School of Medicine, Itabashi-ku, Tokyo, Japan. Electronic address: riku.arai@icloud.com.
Key points
From the paper's conclusion- Serial imaging and surgery demonstrated a perforation-like presentation associated with temporary pacing lead migration, without an open myocardial defect.
Abstract
BACKGROUND: Temporary pacing can cause right ventricular perforation, and lead migration may be associated with myocardial injury mimicking transmural rupture. CASE SUMMARY: A 76-year-old man with sinus node dysfunction underwent temporary pacing, with the lead positioned at the right ventricular septum. On day 4, loss of capture and pericardial effusion developed with lead migration toward the apex. Echocardiography showed no extracardiac protrusion, whereas computed tomography suggested perforation. Because recent anticoagulation increased the potential bleeding consequences if bedside extraction unmasked an occult defect, the lead was removed under direct surgical visualization. Inspection showed apical wall thinning without an open defect or bleeding. Cytology showed reactive mesothelial and inflammatory cells. DISCUSSION: Serial imaging and surgery demonstrated a perforation-like presentation associated with temporary pacing lead migration, without an open myocardial defect. TAKE-HOME MESSAGE: New effusion with pacing dysfunction warrants urgent evaluation and controlled extraction if perforation cannot be excluded.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Delayed Right Atrial Lead Perforation: Discrepancy Between Preoperative Findings and Surgical Inspection.Pacing and clinical electrophysiology : PACE · 2026
- Subacute Microperforation of Right Atrial Pacemaker Lead Presenting with Shock due to Cardiac Tamponade: A Case Report.Clinical practice and cases in emergency medicine · 2026
- Successful management of cardiac tamponade due to right ventricular perforation following temporary pacing wire removal after primary percutaneous coronary intervention at a district general hospital: a case report.European heart journal. Case reports · 2026
- Case Report: Electrode perforation with normal pacemaker parameters leading to hemorrhagic shock: a rare case report of the "penetration-exit-re-entry" phenomenon.Frontiers in medicine · 2026
- Case Report: Double-click on the ventricle: ventricular safety pacing triggered by right atrial lead dislodgement in a patient after right atrial tumor resection.Frontiers in cardiovascular medicine · 2026
- Late intrathoracic migration of a retained temporary epicardial pacing wire associated with middle lobe atelectasis and hemoptysis 15 years after aortic valve replacement: a case report.General thoracic and cardiovascular surgery cases · 2026