First-in-Human Axillary Vein Implantation of a Leadless Pacemaker: Breaking the Access Barrier.
Lydia Messele, William Leon, Niailah Ochai, Gem Pulmano, Uma N Srivatsa
Division of Cardiovascular Medicine, Department of Internal Medicine, University of California-Davis Health, Sacramento, California, USA. Electronic address: lydia.messele@gmail.com. · Division of Cardiovascular Medicine, Department of Internal Medicine, University of California-Davis Health, Sacramento, California, USA.
Key points
From the paper's conclusion- Axillary venous access is a feasible bailout route for LP implantation when femoral and internal jugular access are unavailable.
Abstract
BACKGROUND: Leadless pacemakers (LPs) are conventionally implanted transfemorally, with transjugular and subclavian routes reported as alternatives. Axillary access has not previously been reported. FIH SUMMARY: A 78-year-old woman with chronic sacral wounds became temporary-pacing dependent following an inferior ST-segment elevation myocardial infarction. An LP was selected given infection risk. Because of a right femoral deep venous thrombosis, an abscess at the left femoral site, and restricted neck mobility limiting internal jugular access, she underwent left axillary vein implantation of an LP. DISCUSSION: Axillary access served as a feasible alternative route. A septal position was achieved by rotating the delivery system counterclockwise, opposite the transfemoral direction, while a perpendicular sterile table supported the extracorporeal catheter length. NOVELTY: This is the first reported axillary vein implantation of an LP. TAKE-HOME MESSAGE: Axillary venous access is a feasible bailout route for LP implantation when femoral and internal jugular access are unavailable.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Safety and feasibility of large-bore internal jugular vein access for leadless pacemaker implantation: A systematic review and meta-analysis.The journal of vascular access · 2026
- Dual-Chamber Leadless Pacemaker Implantation Via a Left-Sided Inferior Vena Cava After Pacemaker Extraction.JACC. Case reports · 2026
- Implantation of a Leadless Pacemaker Via a Trans-Iliac Vein Stent Approach: A Case Report.Pacing and clinical electrophysiology : PACE · 2026
- Internal jugular vein approach facilitating second leadless pacemaker implantation after battery depletion.Journal of cardiology cases · 2026
- Feasibility and safety of leadless pacemaker implantation via the right internal jugular vein: primary results of the VAMPIRE multicenter prospective study.Heart rhythm · 2026
- Jugular Journeys: A Novel Approach for AVEIR Leadless Pacemaker ImplantationPacing and Clinical Electrophysiology · 2026