Feasibility and safety of leadless pacemaker implantation via the right internal jugular vein: primary results of the VAMPIRE multicenter prospective study.
Pierre Ollitrault, Gaël Clerici, Andreas Mussigbrodt, Mayane A L Khoury, Karine Bauley, Nathalie Behar, Mouna Ben Kilani, Arnaud Bisson,
Cardiology Department, CHU Caen, Avenue de la Côte de Nacre, 14000, Caen. Electronic address: ollitrault-p@chu-caen.fr. · Cardiology Department, CHU de la Réunion, 212 allée des Topazes, CS 11021, 97400 Saint-Denis.
Key points
From the paper's conclusion- In this prospective multicenter registry of a first-line RIJV approach for LP implantation, feasibility and procedural success were high, and 30-day major complications rate low. These results support the RIJV route as a safe and broadly applicable access strategy for LP implantation. However, dedicated studies are needed to compare safety versus the femoral approach.
Abstract
BACKGROUND: Leadless pacemakers (LPs) are usually implanted through the femoral vein. The right internal jugular vein (RIJV) approach has emerged as an alternative, but evidence remains limited to retrospective single-centre series. OBJECTIVE: To evaluate the feasibility, procedural success, and 30-day safety of a first-line RIJV approach for LP implantation. METHODS: The VAMPIRE study (NCT06455566) is a prospective, investigator-initiated, multicenter registry enrolling consecutive patients with a guideline-directed LP indication across 15 tertiary centres in France (May 2024-May 2026). The primary endpoint was major procedure- or system-related complications within 30 days. RESULTS: Of 294 screened patients, 272 (92.5%) underwent a first-line RIJV approach. The cohort was elderly (median age 79 [73-85] years), predominantly male (63.6%), with high comorbidity burden including atrial fibrillation in 51.5%, chronic kidney disease in 34.9%, and cancer in 22.4%. High-degree atrioventricular block was the primary indication in 84.9%; 65.8% of procedures were performed urgently. Successful LP implantation was achieved in 266 of 272 patients (97.8%; 95% CI 95.3-99.2%), with septal deployment in 94.3%. Median procedure time was 30 [25-40] minutes and median fluoroscopy time 2.5 [1.1-7.0] minutes. Major complications occurred in 6 of 272 patients (2.2%; 95% CI 0.8-4.7%; p<0.001). CONCLUSION: In this prospective multicenter registry of a first-line RIJV approach for LP implantation, feasibility and procedural success were high, and 30-day major complications rate low. These results support the RIJV route as a safe and broadly applicable access strategy for LP implantation. However, dedicated studies are needed to compare safety versus the femoral approach.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Jugular Journeys: A Novel Approach for AVEIR Leadless Pacemaker ImplantationPacing and Clinical Electrophysiology · 2026
- Safety and feasibility of large-bore internal jugular vein access for leadless pacemaker implantation: A systematic review and meta-analysis.The journal of vascular access · 2026
- Atrial and Ventricular Leadless Pacemaker Implantation Using the Internal Jugular Vein: A Multicenter Experience.Pacing and clinical electrophysiology : PACE · 2026
- Jugular versus femoral access for leadless pacemaker implantation: a propensity score–matched multicenter analysisEuropean Heart Journal · 2026
- First-in-Human Axillary Vein Implantation of a Leadless Pacemaker: Breaking the Access Barrier.JACC. Case reports · 2026
- Internal jugular vein approach facilitating second leadless pacemaker implantation after battery depletion.Journal of cardiology cases · 2026