Department of Cardiology, Maastricht University Medical Center (MUMC+), Maastricht, the Netherlands; Cardiovascular Research Institute Maastricht (CARIM), Maastricht University, Maastricht, the Netherlands. Electronic address: m.anass1702@gmail.com. · Department of Cardiology, Maastricht University Medical Center (MUMC+), Maastricht, the Netherlands; Cardiovascular Research Institute Maastricht (CARIM), Maastricht University, Maastricht, the Netherlands.

Case report

Key points

From the paper's conclusion
  • Multimodality fusion imaging-guided electrosurgical leaflet modification using the leaflet laceration with balloon-mediated annihilation to prevent coronary obstruction with radiofrequency needle (LLAMACORN) technique, followed by implantation of a 26-mm self-expanding valve, achieved excellent hemodynamics and preserved coronary perfusion.

Abstract

BACKGROUND: Coronary obstruction is a catastrophic complication of transcatheter valve-in-valve procedures, as the prosthesis frame can form a neo-skirt compromising coronary perfusion. CASE SUMMARY: A 79-year-old woman presented with a degenerated 23-mm balloon-expandable transcatheter valve, heart failure, and reduced left ventricular function. Computed tomography showed high-risk anatomy for right coronary obstruction: valve-to-right coronary artery distance 3.1 mm and valve-to-sinotubular junction distance 1.3 mm. Multimodality fusion imaging-guided electrosurgical leaflet modification using the leaflet laceration with balloon-mediated annihilation to prevent coronary obstruction with radiofrequency needle (LLAMACORN) technique, followed by implantation of a 26-mm self-expanding valve, achieved excellent hemodynamics and preserved coronary perfusion. DISCUSSION: This case shows how computed tomography planning and multimodality imaging enable safe management of complex transcatheter aortic valve-in-transcatheter aortic valve anatomies. When conventional coronary protection is not feasible, LLAMACORN leaflet modification offers a safe alternative to mitigate obstruction risk. TAKE-HOME MESSAGES: Recognize anatomical predictors of coronary obstruction in transcatheter aortic valve-in-transcatheter aortic valve. Understand multimodality imaging's role in guiding leaflet modification. Describe LLAMACORN as a leaflet-modification strategy when a self-expanding valve is preferred.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers