Algorithm-Facilitated Pacemaker-Mediated Tachycardia Mimicking Refractory Septic Shock.
Aradhya Abrol, Japheth Okpebholo, Sarah A Worsnick, Karim Djekidel, Pugazhendhi Vijayaraman
Department of Internal Medicine, Geisinger Wyoming Valley Medical Center, Wilkes-Barre, Pennsylvania, USA. Electronic address: aradhyaabrol98@gmail.com. · Department of Cardiology, Section of Cardiac Electrophysiology, Geisinger Heart Institute, Wilkes-Barre, Pennsylvania, USA.
Key points
From the paper's conclusion- Suspect slow PMT in paced patients with fixed tachycardia near 100-110 beats/min when AV hysteresis is active.
Abstract
BACKGROUND: Algorithms that minimize right ventricular pacing, such as atrioventricular (AV) hysteresis, prolong the AV delay to promote intrinsic conduction. With intact ventriculoatrial (VA) conduction, this can facilitate pacemaker-mediated tachycardia (PMT). CASE SUMMARY: A 62-year-old woman with a dual-chamber pacemaker presented with respiratory failure and hypotension at a paced rate of 110 beats/min, treated for presumed sepsis with escalating vasopressors. Interrogation showed an upper tracking rate of 130 beats/min, AV hysteresis ON, and postventricular atrial refractory period (PVARP) (225 milliseconds) shorter than VA conduction (360 milliseconds), producing a 550-millisecond endless-loop PMT (∼109 beats/min) that escaped device detection. Disabling hysteresis and extending PVARP terminated it, allowing vasopressor weaning. DISCUSSION: When PVARP is shorter than the VA interval, a long AV delay converts retrograde conduction into a slow re-entrant circuit that escapes device diagnostics and mimics sinus tachycardia. TAKE-HOME MESSAGES: Suspect slow PMT in paced patients with fixed tachycardia near 100-110 beats/min when AV hysteresis is active. Disabling hysteresis and extending PVARP beyond VA conduction is diagnostic and lifesaving.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Pseudo-pacemaker syndrome triggered by atrioventricular hysteresis algorithm activation in a patient with left bundle branch area pacing: a case report.Herzschrittmachertherapie & Elektrophysiologie · 2026
- Clues From the Device: Individualized Programming to Solve a Pacing-Facilitated AVNRT.Pacing and clinical electrophysiology : PACE · 2026
- Sub-Acute Re-Elevated Pacing Thresholds After Atrial Leadless Pacemaker Implantation.Journal of cardiovascular electrophysiology · 2026
- Seizure Like Activity During Permanent Pacemaker Implantation: A Case of Suspected Venous Air Embolism With Acute Hypoxia.Clinical case reports · 2026
- Bradycardia-Dependent Loss of Atrial Capture After Single-Chamber Leadless Atrial Pacemaker Implantation.JACC. Case reports · 2026
- Heart-Rate-Dependent Right-to-Left Shunting Through a Patent Foramen Ovale in Severe Right Ventricular Dysfunction: A Case Report.Journal of clinical medicine · 2026