Herzschrittmachertherapie & Elektrophysiologie

Pseudo-pacemaker syndrome triggered by atrioventricular hysteresis algorithm activation in a patient with left bundle branch area pacing: a case report.

Luca Nolte, Corinna Lenz, Max Olav Schrage, Hinrich Schroer

Department of Internal Medicine and Cardiology, BG Klinikum Unfallkrankenhaus Berlin, Warener Str. 7, 12683, Berlin, Germany. · Department of Internal Medicine and Cardiology, BG Klinikum Unfallkrankenhaus Berlin, Warener Str. 7, 12683, Berlin, Germany. Hinrich.Schroer@med.uni-rostock.de.

Case report

Key points

From the paper's conclusion
  • This case report describes a young patient with left bundle branch area pacing (LBBAP) in whom activation of an AV-hysteresis algorithm resulted in excessive prolongation of AV intervals, leading to ineffective AV synchrony and recurrence of symptoms consistent with a pseudo-pacemaker syndrome despite permanent pacemaker implantation, including dizziness and near-syncope.

Abstract

Atrioventricular (AV) hysteresis algorithms are designed to minimize unnecessary ventricular pacing. This case report describes a young patient with left bundle branch area pacing (LBBAP) in whom activation of an AV-hysteresis algorithm resulted in excessive prolongation of AV intervals, leading to ineffective AV synchrony and recurrence of symptoms consistent with a pseudo-pacemaker syndrome despite permanent pacemaker implantation, including dizziness and near-syncope. The case highlights the importance of careful device programming, particularly in patients undergoing conduction system pacing.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers