Clinical case reports
Seizure Like Activity During Permanent Pacemaker Implantation: A Case of Suspected Venous Air Embolism With Acute Hypoxia.
Sanjida Ansari, Murari Dhungana
Shahid Gangalal National Heart Center Kathmandu Nepal.
Key points
From the paper's conclusion- A 78 year old man developed sudden generalized stiffness, seizure‐like activity, and severe hypoxia (oxygen saturation 70%) during lead manipulation without hypotension or arrhythmia.
Abstract
Venous air embolism is a rare complication of pacemaker implantation. A 78 year old man developed sudden generalized stiffness, seizure-like activity, and severe hypoxia (oxygen saturation 70%) during lead manipulation without hypotension or arrhythmia. Prompt recognition, high-flow oxygen, and temporary procedural interruption led to complete recovery without neurological sequelae.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Leadless Atrial Pacemaker Migration Presenting as Gluteal Spasms.JACC. Case reports · 2026
- Case Report: Electrode perforation with normal pacemaker parameters leading to hemorrhagic shock: a rare case report of the "penetration-exit-re-entry" phenomenon.Frontiers in medicine · 2026
- Venous Air/Gas Embolism During Surgical Hysteroscopy: An Underestimated, Potentially Lethal Complication.Cureus · 2026
- Superior vena cava perforation during atrial lead implantation in a patient with repaired tetralogy of Fallot: a case report.European heart journal. Case reports · 2026
- Torque Transmission Failure During Leadless Pacemaker Implantation Caused by Severe Iliofemoral Venous Tortuosity.JACC. Case reports · 2026
- Subacute Microperforation of Right Atrial Pacemaker Lead Presenting with Shock due to Cardiac Tamponade: A Case Report.Clinical practice and cases in emergency medicine · 2026