Cone-Beam CT-Guided SpineJack® vertebral augmentation: A reproducible technique for accurate pedicular access and implant deployment.
Emmanuel Dien, Sylvain Grange, Paul Mondon, Lucas Peraldi, Joris Lavigne, Bastien Chalamet, Mohamed Ilyess Ben-Rejeb, Marion Bouchet,
Department of Interventional Radiology, University Hospital of Nice (CHU Nice), Université Côte d'Azur, Nice, France. Electronic address: dien.e@chu-nice.fr. · Department of Radiology, University Hospital of Saint-Étienne (CHU Saint-Étienne), Saint-Étienne, France.
Key points
From the paper's conclusion- CBCT-guided SpineJack augmentation achieved accurate bilateral implant deployment and substantial immediate vertebral height restoration without major procedure-related complications. These findings support CBCT as a reliable guidance strategy for vertebral augmentation, with demonstrated feasibility in outpatient practice.
Abstract
BACKGROUND: SpineJack® vertebral augmentation requires precise bilateral transpedicular access and accurate implant deployment. This study evaluated the feasibility, safety, and technical performance of cone-beam computed tomography (CBCT)-guided SpineJack augmentation. METHODS: This retrospective single-institution, two-site study included consecutive CBCT-guided procedures performed between May 2021 and April 2026. Technical success required planned bilateral implantation, expansion, and cement augmentation without pedicle breach or implant malposition on final CBCT. Secondary outcomes included immediate vertebral height restoration, complications, procedure duration, radiation exposure, and treatment setting. RESULTS: The analysis included 178 procedures; median age was 45 years (interquartile range [IQR], 27-59), and 158 procedures (88.8%) involved traumatic fractures. Technical success was 100% (178/178), including 64 procedures (36.0%) with pedicle-implant clearance < 0.8 mm. Median absolute vertebral body height gain was 4.1 mm (IQR, 2.7-6.1), and median restoration of baseline height loss reached 77% (IQR, 47-98). All 24 cement leaks (13.5%) were asymptomatic. No neurological, vascular, infectious, or implant-related complications occurred. Median procedure duration was 60 min (IQR, 50-90), and median dose-area product was 33.2 Gy·cm² (IQR, 19.4-53.1). One hundred procedures (56.2%) were performed on an outpatient basis. CONCLUSIONS: CBCT-guided SpineJack augmentation achieved accurate bilateral implant deployment and substantial immediate vertebral height restoration without major procedure-related complications. These findings support CBCT as a reliable guidance strategy for vertebral augmentation, with demonstrated feasibility in outpatient practice.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- The accuracy of pedicle screw placement in thoracolumbar fractures: A comparative study of automated cone-beam CT-based navigation and conventional fluoroscopy.Brain & spine · 2026
- Fully percutaneous, navigation-assisted cervical pedicle screw instrumentation: technical note and a report of safety and accuracy.European spine journal : official publication of the European Spine Society, the European Spinal Deformity Society, and the European Section of the Cervical Spine Research Society · 2026
- Long-Term Effect of SpineJack in the Treatment of Traumatic and Osteoporotic Vertebral Fractures.Journal of clinical medicine · 2026
- Anatomical and biological reconstruction of traumatic complete burst fractures of the thoracolumbar vertebrae using pedicle fixation combined with stent-armed kyphoplasty and intracorporeal bone grafting: technique rationale, 7-year experience and clinical, imaging, and histological outcomes.Journal of spine surgery (Hong Kong) · 2026
- Intraoperative three-dimensional imaging-guided vertebral body stenting for cleft-associated very severe osteoporotic vertebral compression fractures: a surgical technique case series.Journal of spine surgery (Hong Kong) · 2026
- Procedural Complications and Implant Cutout of the SpineJack®-System in Thoracolumbar Vertebral Fractures: A Retrospective Analysis of 204 Implants from a Level-1 Trauma Center.Global spine journal · 2026