Procedural Complications and Implant Cutout of the SpineJack®-System in Thoracolumbar Vertebral Fractures: A Retrospective Analysis of 204 Implants from a Level-1 Trauma Center.
Julian Scherer, Magdalena Karner, Georg Osterhoff, Kai Sprengel, Hans-Christoph Pape
Department of Traumatology, University Hospital of Zurich, Zurich, Switzerland. · Department of Orthopaedics, Schulthess Klinik Zurich, Zurich, Switzerland.
Key points
From the paper's conclusion- This study determined the incidence of procedural complications, with a focus on cutout, and identified associated factors.MethodsAll patients treated with stand-alone SpineJack® stent-kyphoplasty at a level-1 trauma center (November 2014-December 2022) were retrospectively reviewed, excluding malignant fractures.
Abstract
Study DesignRetrospective cohort study.ObjectivesImplant-related complications of SpineJack® stent-kyphoplasty, particularly endplate cutout, remain poorly characterized. This study determined the incidence of procedural complications, with a focus on cutout, and identified associated factors.MethodsAll patients treated with stand-alone SpineJack® stent-kyphoplasty at a level-1 trauma center (November 2014-December 2022) were retrospectively reviewed, excluding malignant fractures. In total, 204 implants were analyzed (129 OVFs, 75 VCFs). Intraoperative complications and postoperative endplate cutout, assessed on imaging after mobilization, were analyzed by fracture type, AO/OF classification, region and vertebral level.ResultsMean age was 64.9 years (SD 17.5); OVF patients were significantly older than VCF patients (74.3 vs 48.8 years, p<0.001). Vascular cement leakage occurred in 2 cases (1.0%); no technical failures or conversions were recorded. Cement leakage occurred in 18 implants (8.8%) and cutout in 16 (7.8%), neither differing by fracture type. Cutout incidence differed significantly across OF grades (p=0.018), though not after excluding the small OF1 subgroup (p=0.211) or in multivariable analysis (OR 1.03, p=0.918), suggesting limited power; cutout was not correlated with cement leakage or vertebral level.ConclusionsStand-alone SpineJack® stent-kyphoplasty is associated with infrequent vascular and cement-related complications. Endplate cutout (7.8%) was numerically more frequent in severely collapsed osteoporotic (OF4) fractures, but this association was not robust to sensitivity or multivariable analysis, consistent with limited power rather than a true absence of effect; no cutout event in this cohort had a demonstrated clinical consequence. Prospective, adequately powered study is needed to confirm this association.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Intervertebral Disc Abnormalities on MRI in Thoracolumbar High-Energy and Fragility Fractures Treated With the SpineJack®-System: A Secondary Analysis of a Single-Centre Cohort.Global spine journal · 2026
- Clinical and radiographic outcomes of SpineJack mechanical kyphoplasty versus conservative management in traumatic thoraco-lumbar compression fractures.The surgeon : journal of the Royal Colleges of Surgeons of Edinburgh and Ireland · 2026
- Risk factors associated with bone cement leakage following percutaneous kyphoplasty: a retrospective analysisEuropean journal of medical research · 2025
- Development and internal validation of a prediction model for new vertebral compression fractures after percutaneous kyphoplasty: a retrospective studyBMC Musculoskeletal Disorders · 2026
- Endplate-to-endplate cement contact and 24-month outcomes after unilateral percutaneous vertebroplasty for osteoporotic vertebral compression fractures: a single-center retrospective cohort study.Frontiers in surgery · 2026
- Long-Term Effect of SpineJack in the Treatment of Traumatic and Osteoporotic Vertebral Fractures.Journal of clinical medicine · 2026