Quantitative assessment of vertebral bone marrow adipose tissue using photon-counting CT compared with MRI.
Dingbo Shu, Jianping Wang, Ding Liang, Li Zhao, Zhenhua Zhao
Department of Radiology, Shaoxing People's Hospital, Shaoxing, China; School of Medicine, Shaoxing University, Shaoxing, China. · Department of Radiology, Shaoxing People's Hospital, Shaoxing, China; School of Medicine, Shaoxing University, Shaoxing, China. Electronic address: zhao2075@163.com.
Key points
From the paper's conclusion- PCCT-derived VNCa CTFF showed strong correlation with MRI PDFF, and linear calibration reduced the mean difference between the two measurements, although the limits of agreement remained relatively wide. Standardized 70 keV CT values were also correlated with MRI PDFF, but should be interpreted cautiously as attenuation-based surrogates influenced by both trabecular mineral content and marrow fat.
Abstract
RATIONALE AND OBJECTIVES: To evaluate the feasibility and accuracy of photon-counting CT (PCCT)-derived parameters for quantifying vertebral bone marrow adipose tissue (BMAT), using MRI proton density fat fraction (PDFF) as the reference standard. MATERIALS AND METHODS: Adult participants with low back pain were prospectively recruited to undergo both PCCT and MRI. PCCT-derived parameters, including virtual non-calcium (VNCa) CT values, VNCa CT fat fraction (CTFF), and standardized 70 keV CT values, were obtained. MRI PDFF maps were generated using a 6-echo q-Dixon sequence. Correlation and agreement between PCCT-derived parameters and MRI PDFF were assessed at both vertebral and subject levels. RESULTS: A total of 73 participants were included, and 339 vertebral bodies were analyzed. VNCa CTFF showed strong positive correlation with MRI PDFF at both vertebral (r = 0.863) and subject (r = 0.895) levels. In contrast, VNCa CT values (r = -0.858 to -0.888) and standardized 70 keV CT values (r = -0.791 to -0.833) showed significant negative correlations with MRI PDFF. Bland-Altman plots showed small mean differences but relatively wide limits of agreement between calibrated VNCa CTFF and MRI PDFF. CONCLUSIONS: PCCT-derived VNCa CTFF showed strong correlation with MRI PDFF, and linear calibration reduced the mean difference between the two measurements, although the limits of agreement remained relatively wide. Standardized 70 keV CT values were also correlated with MRI PDFF, but should be interpreted cautiously as attenuation-based surrogates influenced by both trabecular mineral content and marrow fat.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Multi-Organ Fat Fraction Quantification with Photon-Counting CT: Determinants, Regional Heterogeneity, and Agreement Between Spectral Fat Maps and Attenuation-Derived Fat Fraction in 270 Patients.Journal of clinical medicine · 2026
- Robust visualization of the right adrenal vein using photon-counting CT across all BMI groups.Abdominal radiology (New York) · 2026
- Robust Liver Steatosis Quantification Using Photon-Counting CT in Comparison to PDFF: Validation across Protocols and Cohorts.Academic radiology · 2026
- Uniform quantification of hepatic fat fraction across body sizes using photon-counting CT.European journal of radiology · 2026
- Utility of photon-counting CT localizer radiograph in opportunistic bone densitometry.Journal of medical imaging (Bellingham, Wash.) · 2026
- Reconstruction-kernel Dependence of Femoral Bone Marrow Attenuation in Whole-body Dual-source Photon-counting CT of Patients with Multiple Myeloma.Academic radiology · 2026