Robust visualization of the right adrenal vein using photon-counting CT across all BMI groups.
Meng Xu, Lan Liu, Rao Song, Haiping Zhang, Xiaojing He, Xiaojing Zhang, Chuanjun Zhong, Dajing Guo,
The Second Affiliated Hospital of Chongqing Medical University, Chongqing, China. · Siemens (China), Shanghai, China.
Key points
From the paper's conclusion- PCCT provides favorable RAV visualization across the investigated BMI groups, with low-keV imaging and smoother reconstruction kernels associated with improved visualization quality.
Abstract
OBJECTIVE: To evaluate the robust visualization of photon-counting CT (PCCT) for right adrenal vein (RAV) across body mass index (BMI) groups with respect to virtual monochromatic imaging (VMI) energy levels and reconstruction kernels. METHODS: A total of 302 patients who underwent contrast-enhanced abdominal PCCT were enrolled and stratified into non-obese, overweight, and obese groups based on Asian-specific BMI criteria. VMIs were reconstructed at 40, 55, and 70 keV with three vascular‑specific reconstruction kernels (Bv40, Bv48, Bv60). For quantitative evaluation, two radiologists assessed image metrics, including CT attenuation, image noise (standard deviation [SD]), signal-to-noise ratio (SNR), and contrast-to-noise ratio (CNR). Separately, another pair of radiologists independently evaluated qualitative image quality (encompassing both overall image quality and RAV visualization quality) using a 5-point Likert scale. RESULTS: Quantitative evaluation showed that VMI energy level and reconstruction kernel were associated with changes in quantitative image quality metrics, with significant interactions between energy level and reconstruction kernel for all four quantitative metrics (all P < 0.001). Exploratory analyses suggested BMI-related interactions in RAV attenuation and subjective visualization quality. Lower keV increased RAV CT attenuation and CNR, with the highest values at 40 keV across all BMI groups. Qualitative evaluation demonstrated a 100% RAV visualization rate across all reconstruction settings and BMI groups, with median visualization scores ranging from 3.0 to 5.0. Inter-reader agreement was excellent (weighted κ = 0.82). CONCLUSION: PCCT provides favorable RAV visualization across the investigated BMI groups, with low-keV imaging and smoother reconstruction kernels associated with improved visualization quality.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Intraindividual Comparison of Radiation Dose and Image Quality: Photon-counting versus Energy-integrating Detector CT in Obesity.Radiology · 2026
- Optimization of reconstruction strategies for head and neck computed tomography angiography using photon-counting computed tomography: A comparative study with energy-integrating detector computed tomography.European journal of radiology · 2026
- Quantitative assessment of vertebral bone marrow adipose tissue using photon-counting CT compared with MRI.European journal of radiology · 2026
- Feasibility of low concentration iodinated contrast medium in portal venous abdominal imaging: performance of photon-counting CT virtual monoenergetic imaging compared to conventional EID-CT.European journal of radiology · 2026
- Plaque-Adapted Virtual Monoenergetic Image Selection for Carotid Plaque Visualization Using Photon-Counting CT.Diagnostics (Basel, Switzerland) · 2026
- Dual-Source Photon-Counting CT for Thoracic Arteries: Enhancing Image Quality and Diagnostic Assessability with Low-Energy Virtual Monoenergetic Imaging.Diagnostics (Basel, Switzerland) · 2026