Parameter-guided correction of B1 inhomogeneity in 3D VFA T1 mapping: A feasibility study using MRI simulation.
Nagomi Fukuda, Yuki Kanazawa, Akihiro Haga, Ryuji Ohshiro, Shun Kitano, Hiroaki Hayashi, Motoharu Sasaki, Masafumi Harada
Department of Radiological Technology, Tokushima University Hospital, Tokushima, Japan. Electronic address: fukuda.nagomi@tokushima-u.ac.jp. · Department of Medical Imaging Physics, Faculty of Life Sciences, Kumamoto University, Kumamoto, Japan; Institute of Biomedical Sciences, Tokushima University Graduate School, Tokushima, Japan. Electronic address: ykz@kumamoto-u.ac.jp.
Key points
From the paper's conclusion- These findings support the feasibility of parameter-guided correction under the tested simulations but show dependence on predefined spatial parameters. The framework may avoid a dedicated B1-mapping sequence when the required α/2α images are available; however, it does not estimate a subject-specific B1+ field and requires prospective validation.
Abstract
BACKGROUND: Three-dimensional variable flip angle (VFA) T1 mapping provides rapid quantitative imaging but is highly sensitive to transmit B1 inhomogeneity. Although B1 mapping can reduce this bias, it requires additional scans and is not always feasible in clinical protocols. PURPOSE: To evaluate the feasibility and model sensitivity of a parameter-guided B1 correction framework for VFA T1 mapping. METHODS: Bloch-based simulations generated multi-FA spoiled gradient-echo (SPGR) images in cylindrical and digital brain phantoms. Pseudo-B1-related signal nonuniformity was imposed by multiplying a normalized spatial weighting voxelwise after Bloch signal generation. The empirical correction coefficient combined a double-flip-angle (DFA)-inspired signal-ratio term with a predefined Gaussian spatial prior. T1 was estimated using non-negative linear least-squares fitting. Separately, an in vivo-derived B1+ pattern was applied to simulated SPGR data, and the parameter-guided correction was evaluated. RESULTS: In the cylindrical phantom, corrected in-plane T1 bias was < 5% under centered conditions and < 7% (TR = 10 ms) or < 3% (TR = 100 ms) at slice edges. Under a shifted-center condition, residual bias reached 119.9%, indicating sensitivity to spatial model mismatch. In the digital brain phantom, local biases after correction were within 5%. Using an in vivo-derived B1+ pattern, the parameter-guided correction qualitatively reduced T1 variation compared with no correction. CONCLUSION: These findings support the feasibility of parameter-guided correction under the tested simulations but show dependence on predefined spatial parameters. The framework may avoid a dedicated B1-mapping sequence when the required α/2α images are available; however, it does not estimate a subject-specific B1+ field and requires prospective validation.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Incomplete Spoiling Correction for 3D T 1 Mapping.Magnetic resonance in medicine · 2026
- Variable flip angle T1 mapping for quantitative liver imaging at 0.55 T.Magma (New York, N.Y.) · 2026
- A correction to the bandwidth difference method for assessment of magnetic resonance static field homogeneity.Medical physics · 2026
- High Dynamic Range Turbo-FLASH B1 Mapping for Characterizing Multi-Transmit Multi-Receive RF Arrays, Validated at 7 T and 11.7 T.Magnetic resonance in medicine · 2026
- Fat Fraction Mapping in the Presence of Magnetic Field Inhomogeneities With Phase‐Cycled bSSFP at 3 TNMR in Biomedicine · 2026
- Self-Correction of B0 Drift and Trajectory Error in Alternating Unbalanced SSFP-FID and SSFP-ECHO (AUSFIDE)-Based 3D Transverse Relaxometry of the Human Brain.Magnetic resonance in medicine · 2026