Magma (New York, N.Y.)Open accessSegmentation

Variable flip angle T1 mapping for quantitative liver imaging at 0.55 T.

Majd Helo, Marcel Dominik Nickel, Thomas Küstner

Medical Image and Data Analysis (MIDAS.Lab), Department of Diagnostic and Interventional Radiology, University of Tuebingen, Tuebingen, Germany. majd.helo@outlook.com. · Research and Clinical Translation, Magnetic Resonance, Siemens Healthineers AG, Erlangen, Germany.

Key points

From the paper's conclusion
  • At 0.55 T, RF spoiling and reconstruction strategies are the main determinants of quantitative accuracy, while transmit field inhomogeneity plays a minor role. Combined RF spoiling correction and DL reconstruction enables robust VFA T1 mapping without B1+ correction. Reduced low-field T1 values improve VFA model conditioning and reduce sensitivity to flip angle deviations, supporting VFA-based T1 mapping for low-field MRI.

Abstract

PURPOSE: Investigating the feasibility of variable flip angle (VFA) T1 mapping for quantitative liver imaging at 0.55 T and evaluating the impact of RF spoiling correction and deep learning-based (DL) reconstruction on accuracy. METHODS: A VFA protocol was developed and combined with DL reconstruction to improve quantitative robustness at 0.55 T. Phantom experiments validated T1 estimates against multi-TI inversion-recovery spin-echo reference measurements. Bloch simulations characterized RF spoiling-induced bias and derived 0.55 T-specific calibration coefficients. Twelve healthy subjects were scanned to assess field-strength-dependent effects of B1+ inhomogeneity. RESULTS: In vivo, DL reconstruction reduced noise-related variability of voxel-wise hepatic T1 estimates compared to conventional reconstructions in the same subject. At 0.55 T, B1+ maps showed the highest transmit homogeneity, with liver flip angle variation of 98.2 ± 2.9%, compared with 94.6 ± 7.4% at 1.5 T and 100.7 ± 9.8% at 3 T. Accordingly, B1+ correction had minimal impact on hepatic T1 distributions at 0.55 T, whereas more pronounced effects were observed at higher field strengths. Phantom experiments demonstrated improved agreement of VFA-derived T1 values with the IR-SE reference after 0.55 T-specific RF spoiling correction, supporting the accuracy of the corrected VFA approach. Bland-Altman analysis confirmed that RF spoiling correction had a greater impact on phantom T1 accuracy than B1+ correction. CONCLUSION: At 0.55 T, RF spoiling and reconstruction strategies are the main determinants of quantitative accuracy, while transmit field inhomogeneity plays a minor role. Combined RF spoiling correction and DL reconstruction enables robust VFA T1 mapping without B1+ correction. Reduced low-field T1 values improve VFA model conditioning and reduce sensitivity to flip angle deviations, supporting VFA-based T1 mapping for low-field MRI.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers