Variable flip angle T1 mapping for quantitative liver imaging at 0.55 T.
Majd Helo, Marcel Dominik Nickel, Thomas Küstner
Medical Image and Data Analysis (MIDAS.Lab), Department of Diagnostic and Interventional Radiology, University of Tuebingen, Tuebingen, Germany. majd.helo@outlook.com. · Research and Clinical Translation, Magnetic Resonance, Siemens Healthineers AG, Erlangen, Germany.
Key points
From the paper's conclusion- At 0.55 T, RF spoiling and reconstruction strategies are the main determinants of quantitative accuracy, while transmit field inhomogeneity plays a minor role. Combined RF spoiling correction and DL reconstruction enables robust VFA T1 mapping without B1+ correction. Reduced low-field T1 values improve VFA model conditioning and reduce sensitivity to flip angle deviations, supporting VFA-based T1 mapping for low-field MRI.
Abstract
PURPOSE: Investigating the feasibility of variable flip angle (VFA) T1 mapping for quantitative liver imaging at 0.55 T and evaluating the impact of RF spoiling correction and deep learning-based (DL) reconstruction on accuracy. METHODS: A VFA protocol was developed and combined with DL reconstruction to improve quantitative robustness at 0.55 T. Phantom experiments validated T1 estimates against multi-TI inversion-recovery spin-echo reference measurements. Bloch simulations characterized RF spoiling-induced bias and derived 0.55 T-specific calibration coefficients. Twelve healthy subjects were scanned to assess field-strength-dependent effects of B1+ inhomogeneity. RESULTS: In vivo, DL reconstruction reduced noise-related variability of voxel-wise hepatic T1 estimates compared to conventional reconstructions in the same subject. At 0.55 T, B1+ maps showed the highest transmit homogeneity, with liver flip angle variation of 98.2 ± 2.9%, compared with 94.6 ± 7.4% at 1.5 T and 100.7 ± 9.8% at 3 T. Accordingly, B1+ correction had minimal impact on hepatic T1 distributions at 0.55 T, whereas more pronounced effects were observed at higher field strengths. Phantom experiments demonstrated improved agreement of VFA-derived T1 values with the IR-SE reference after 0.55 T-specific RF spoiling correction, supporting the accuracy of the corrected VFA approach. Bland-Altman analysis confirmed that RF spoiling correction had a greater impact on phantom T1 accuracy than B1+ correction. CONCLUSION: At 0.55 T, RF spoiling and reconstruction strategies are the main determinants of quantitative accuracy, while transmit field inhomogeneity plays a minor role. Combined RF spoiling correction and DL reconstruction enables robust VFA T1 mapping without B1+ correction. Reduced low-field T1 values improve VFA model conditioning and reduce sensitivity to flip angle deviations, supporting VFA-based T1 mapping for low-field MRI.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Parameter-guided correction of B1 inhomogeneity in 3D VFA T1 mapping: A feasibility study using MRI simulation.Physica medica : PM : an international journal devoted to the applications of physics to medicine and biology : official journal of the Italian Association of Biomedical Physics (AIFB) · 2026
- Free-Breathing PDFF and R2* Quantification at 0.55T: Reduced Respiratory Motion Sensitivity and a Time-Resolved 4D MRI ApproacharXiv · 2026
- 0.55 T Magnetic Resonance Imaging in the Abdomen and Pelvis: Opportunities and Challenges.Journal of computer assisted tomography · 2026
- Water-T1, PDFF, and T2* mapping of the liver using chemical shift-encoded MOLLI.Abdominal radiology (New York) · 2026
- Fat Fraction Mapping in the Presence of Magnetic Field Inhomogeneities With Phase‐Cycled bSSFP at 3 TNMR in Biomedicine · 2026
- Single-shot T2-Weighted Turbo Spin-echo Imaging of the Liver Improves Lesion Characterization Compared with Free-breathing Acquisition.Academic radiology · 2026