Self-Correction of B0 Drift and Trajectory Error in Alternating Unbalanced SSFP-FID and SSFP-ECHO (AUSFIDE)-Based 3D Transverse Relaxometry of the Human Brain.
Eunseo Bae, Sung Suk Oh, Dongyeob Han, Kathryn M Jaroszynski, Felix W Wehrli, Hyunyeol Lee
School of Electronic and Electrical Engineering & IEDT, Kyungpook National University, Daegu, Republic of Korea. · Medical Device Development Center, Daegu-Gyeongbuk Medical Innovation Foundation (K-MEDI Hub), Daegu, Republic of Korea.
Key points
From the paper's conclusion- The self-corrected AUSFIDE method is able to compensate for errors from B0 drift and k-space trajectory mismatch, and thus is a potentially practical and reliable means for 3D transverse relaxometry.
Abstract
PURPOSE: The alternating unbalanced SSFP (AUSFIDE) pulse sequence, which consists of multi-echo trains of SSFP-FID and SSFP-ECHO modules, enables rapid estimation of R2 and R2', but is prone to errors from temporal drift of the main magnetic field (B0) and k-space trajectory mismatch. Here, we propose a self-correcting AUSFIDE method for joint correction of these two sources of error without requiring additional scans. METHODS: AUSFIDE signals were modeled in the presence of B0 drift and k-space shifts. To estimate and correct for B0 drift over the entire scan, phase navigators without spatial encoding were inserted before and after each multi-echo train. In addition, the intrinsic phase symmetry between SSFP-FID and SSFP-ECHO signals was exploited, enabling self-calibration of the k-space sampling trajectory. Numerical simulations and phantom experiments were performed for proof-of-concept validation, followed by in vivo brain imaging of 10 healthy subjects on two 3 T scanners. Measured parameters were analyzed in six brain regions, and test-retest reproducibility was evaluated on the two MRI platforms. RESULTS: Simulations and phantom imaging consistently demonstrated that B0 drift and trajectory errors resulted in image distortions and heavily overestimated R2'. Experiments further revealed that these artifacts and quantification errors were substantially reduced by the proposed self-correction method. Inter-scan agreement of the quantified parameters was also improved after correction on both scanners. CONCLUSION: The self-corrected AUSFIDE method is able to compensate for errors from B0 drift and k-space trajectory mismatch, and thus is a potentially practical and reliable means for 3D transverse relaxometry.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Fat Fraction Mapping in the Presence of Magnetic Field Inhomogeneities With Phase‐Cycled bSSFP at 3 TNMR in Biomedicine · 2026
- Diffusion broadening of the point spread function in steady-state MRIarXiv · 2026
- Servo Navigation and Phase Equalization Enhanced by Run-Time Stabilization (PEERS) for 3D EPI Time Series.Magnetic resonance in medicine · 2026
- High-spatial resolution, high-quality white matter tractography using DTI with dynamic slice-by-slice B0 shimming in a head-only high-gradient performance 3T MRI scannerarXiv · 2026
- Real-Time Correction of Respiration-Induced B 0 Changes in Oscillating Steady State Imaging (OSSI) fMRI.Magnetic resonance in medicine · 2026
- Motion‐ and Field‐Robust Mesoscopic Whole‐Brain T2*‐Weighted Imaging at 7 and 11.7 T Using Servo NavigationMagnetic Resonance in Medicine · 2026