Quantifying three-dimensional displacement of cardiac implantable electronic devices and implications for safety margins in radiotherapy.
Takahiro Kubota, Ryota Onizuka, Kazuki Kajiwara, Tatsunori Saho, Koichiro Muraki
Department of Radiological Technology, Kokura Memorial Hospital, Kitakyushu, 802-8555, Japan. 1230bota@gmail.com. · Department of Radiological Technology, Kokura Memorial Hospital, Kitakyushu, 802-8555, Japan.
Key points
From the paper's conclusion- This study aimed to quantitatively evaluate the three-dimensional (3D) displacement of cardiac implantable electronic devices (CIEDs) during radiotherapy and derive safety margins.
Abstract
This study aimed to quantitatively evaluate the three-dimensional (3D) displacement of cardiac implantable electronic devices (CIEDs) during radiotherapy and derive safety margins. Sixteen treatment courses (13 patients, 262 fractions) were retrospectively analyzed. Daily device displacement was quantified using orthogonal 2D kV images relative to clinical bone-matching. To evaluate manual registration uncertainty, the intraclass correlation coefficient and minimal detectable change (MDC) were calculated. To assess the impact of patient positioning, displacements were compared between arm-up and arm-down groups. Planning organ at risk volume (PRV) margins were derived from systematic and random errors. A maximum 3D displacement of 24 mm was observed. Manual registration demonstrated good to excellent reproducibility for the primary translational axes. The maximum MDC was 3.6 mm for the 3D displacement vector. While overall displacement distributions showed no significant differences between arm positions, extreme medial displacement significantly increased in the arm-up group (p < 0.05). The calculated PRV margins reached a maximum of 7.3 mm in the longitudinal direction for the arm-up group. Throughout the radiotherapy course, CIEDs can undergo displacement independently of the skeletal structure. The PRV margins derived here provide a quantitative basis for pre-treatment dose evaluation. The observed range of displacement can inform treatment planning and supports verification of the device position at each fraction via image-guided radiotherapy.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Minimum recommended distances between cardiac implantable electronic devices and the applicator surface in low-energy intraoperative radiotherapy of the breast: an in vivo dosimetric study.Physica medica : PM : an international journal devoted to the applications of physics to medicine and biology : official journal of the Italian Association of Biomedical Physics (AIFB) · 2026
- Effect of radiotherapy on cardiac implantable electronic devices - evaluation of potential explanatory variables in more than a decade.Radiation oncology (London, England) · 2026
- Systematic review of physical and dosimetric criteria for managing CIEDs in radiotherapy: A medical physicist's perspective.Journal of applied clinical medical physics · 2026
- Cardiac implantable electronic devices and radiation therapy: long-term outcomesEuropean Heart Journal · 2026
- A 3D CBCT-Based Approach to Assessing Effective Radiation Penetration Depth in Intraoperative Electron Radiotherapy.Diagnostics (Basel, Switzerland) · 2026
- Cardiac segmentation for ventricular tachycardia radioablation and radiotherapy patients with cardiac implants.ArXiv · 2026