A liquid fiducial marker for image-guided proton beam therapy of abdominal tumors.
Sarah Stefanowicz, Jonas R Henriksen, Rasmus Jølck, Stefan Menkel, Michael W Mortensen, Esther G C Troost
OncoRay - National Center for Radiation Research in Oncology, Faculty of Medicine and University Hospital Carl Gustav Carus, Technische Universität Dresden, Helmholtz-Zentrum Dresden - Rossendorf, Dresden, Germany; Department of Radiotherapy and Radiation Oncology, Faculty of Medicine and University Hospital Carl Gustav Carus, Technische Universität Dresden, Dresden, Germany; Helmholtz-Zentrum Dresden - Rossendorf, Institute of Radiooncology - OncoRay, Dresden, Germany; Technical University of Munich, TUM School of Medicine and Health, Department of Radiation Oncology, TUM University Hospital, Munich, Germany. Electronic address: sarah.stefanowicz@mri.tum.de. · Nanovi A/S, Development & Supply, Diplomvej 378, 2800 Kgs. Lyngby, Denmark; Department of Health Technology, Section for Cell and Drug Technologies, Technical University of Denmark, Building 423, DK-2800 Kgs. Lyngby, Denmark. Electronic address: jhen@dtu.dk.
Key points
From the paper's conclusion- The liquid fiducial marker demonstrated comparable visibility and chemical stability with reduced artifact generation and dose perturbation compared to solid markers, indicating superiority for IGPT of abdominal tumors. Optimization of the liquid marker composition may be considered.
Abstract
PURPOSE: This study compared a liquid to solid fiducial markers regarding visibility, artifact characteristics on X-ray imaging and dose perturbation for image-guided proton therapy (IGPT) of abdominal tumors. Additionally, the radio-opaque composition of the liquid fiducial marker was optimized, and the chemical stability after proton beam irradiation was analyzed. MATERIAL AND METHODS: The fiducial markers inserted into an anthropomorphic abdominal phantom were exposed to X-ray, single-energy computed tomography (SECT), dual-energy CT (DECT) and cone-beam CT (CBCT) to determine contrast-to-noise ratios (CNR) and artifacts. Proton dose perturbation estimations were performed using a Monte Carlo dose algorithm. Chemical stability following proton beam irradiation was assessed with biophysical and optical methods. RESULTS: The CNR of the liquid marker with a volume ≥ 100 μl (CNR ≤ 2.69) on 2D X-ray images was superior to the solid markers (CNR ≤ 1.72). The artifact areas (A) and severity in the CT modalities were more pronounced for the solid (e.g. , SECT: A ≤ 279.4 mm3) than for the liquid fiducial marker (A ≤ 30.5 mm3). Dose mitigation of ≥ 50% in a proton beam was calculated for solid fiducial markers, while the liquid marker less influenced the dose distribution. Increasing radio-opaque content in the liquid prototypes improved visibility (CNR ≤ 2.97) without significantly increasing artifact volume (SECT: A ≤ 40.5 mm3) or dose perturbation. No degradation of the liquid fiducial marker's chemical composition was observed. CONCLUSION: The liquid fiducial marker demonstrated comparable visibility and chemical stability with reduced artifact generation and dose perturbation compared to solid markers, indicating superiority for IGPT of abdominal tumors. Optimization of the liquid marker composition may be considered.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Endoscopic ultrasound-guided fiducial marker placement for image-guided proton beam therapy in pancreatic cancer: implications for positional accuracy and position management.Journal of radiation research · 2026
- Fabrication, testing, and clinical validation of an epoxy resin-based range shifter for pencil beam scanning proton therapy.Biomedical physics & engineering express · 2026
- Image quality and artefact reduction for tantalum clips in ocular proton therapy using photon-counting CT.European radiology experimental · 2026
- Automated system for MCNP input generation and ICRP 110 phantom adaptation in 125I prostate brachytherapy with edema.Applied radiation and isotopes : including data, instrumentation and methods for use in agriculture, industry and medicine · 2026
- Semi-Quantitative In Vivo Imaging of Intratumorally Injected Iodine-125-Labeled Nanoparticles Using a Benchtop 2D Multi-Pinhole X-ray System.Molecular imaging and biology · 2026
- Pulmonary embolism following spacer hydrogel intravasation: A case report of a rare rectal spacer complication in prostate cancer radiotherapy.Central European journal of urology · 2026